BMC Ophthalmol. 2022 Jul 5;22(1):294. doi: 10.1186/s12886-022-02499-4.
The central hole of the Visian Implantable Collamer Lens (ICL V4c) provides a reference to observe its tilt or decentration. This study aimed to investigate the tilt and decentration effects of ICL V4c on visual quality after implantation.
A total of 135 eyes from 69 patients who underwent ICL V4c implantation were included in this study. Evaluation of uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), and aberrations were performed 6-months postoperatively. The anterior segment parameters were collected using CASIA2 anterior segment-optical coherence tomography, tilt and decentration of ICL V4c were analyzed using MATLAB software. All patients received questionnaires to investigate the visual quality postoperatively.
The safety and effectiveness were 1.18 ± 0.17 and 1.11 ± 0.18, respectively. No significant changes were observed regarding higher-order and spherical aberrations after the operation; however, coma and trefoil significantly increased compared to preoperative values. The average total decentration and tilt was 0.21 ± 0.12 mm and 2.54 ± 1.00°, respectively. Horizontal, vertical, and total values of tilt and decentration were not significantly associated with postoperative CDVA, UDVA, and aberrations. The most common visual symptom was halo, and 97.04% of patients had a satisfaction score ≥ 8. The total or horizontal tilt was significantly positively correlated with the frequency, severity, and bothersome scores from the questionnaires.
ICL V4c implantation can obtain high visual quality and patient satisfaction. Although the degree of tilt and decentration after ICL V4c implantation was small, a positive effect on subjective visual quality was observed.
Visian 可植入式 Collamer 透镜(ICL V4c)的中央孔提供了观察其倾斜或偏心的参考。本研究旨在探讨 ICL V4c 植入后对视觉质量的倾斜和偏心影响。
本研究共纳入 69 例 135 只眼行 ICL V4c 植入术患者。术后 6 个月评估未矫正远视力(UDVA)、矫正远视力(CDVA)和像差。使用 CASIA2 眼前节光学相干断层扫描采集眼前节参数,使用 MATLAB 软件分析 ICL V4c 的倾斜和偏心。所有患者均接受问卷调查以评估术后视觉质量。
安全性和有效性分别为 1.18±0.17 和 1.11±0.18。术后高阶像差和球差无明显变化,但与术前相比,彗差和三叶草像差显著增加。平均总偏心和倾斜分别为 0.21±0.12mm 和 2.54±1.00°。倾斜和偏心的水平、垂直和总值与术后 CDVA、UDVA 和像差均无显著相关性。最常见的视觉症状是光晕,97.04%的患者满意度评分≥8。总或水平倾斜与问卷调查的频率、严重程度和困扰评分呈显著正相关。
ICL V4c 植入可获得较高的视觉质量和患者满意度。尽管 ICL V4c 植入后倾斜和偏心程度较小,但对主观视觉质量有积极影响。