Buehl Wolf, Stojanac Danijela, Sacu Stefan, Drexler Wolfgang, Findl Oliver
Department of Ophthalmology, Medical University of Vienna, Währinger Gürtel 18-20, A-1090 Vienna, Austria.
Am J Ophthalmol. 2006 Jan;141(1):7-12. doi: 10.1016/j.ajo.2005.08.048.
To compare three different methods of measuring corneal thickness (CT) and anterior chamber depth (ACD).
Prospective clinical trial (Medical University of Vienna, Austria).
Central CT (CCT), CT at four peripheral points, and central ACD were measured in 88 eyes of 44 healthy subjects with the Pentacam (rotating Scheimpflug camera; Oculus, Wetzlar, Germany), Orbscan I (scanning-slit topography system; Orbtek Inc, Salt Lake City, Utah, USA), and AC-Master (partial coherence interferometry; Zeiss Meditec, Jena, Germany), and the results were compared.
The upper (lower) limits of agreement for CCT measurements were 7.9 (-22.2) microm between AC-Master and Pentacam, 17.6 (-32.5) microm between AC-Master and Orbscan, and 25.2 (-25.9) microm between Pentacam and Orbscan. Correlation was high between all three methods (r = 0.94 to 0.97). The upper and lower limits of agreement for ACD were 0.174 (-0.251) mm between AC-Master and Pentacam, 0.406 (-0.004) mm between AC-Master and Orbscan, and 0.384 (0.095) mm between Pentacam and Orbscan. Correlation was high between the three methods (r = 0.96 between Orbscan and Pentacam; others 0.92). Correlation was lower for the CT measurements at the four peripheral points.
The CCT and ACD values obtained by Pentacam, Orbscan, and AC-Master measurements correlated well and showed few outliers. The two new systems (Pentacam, AC-Master) provide a reliable, easy-to-use, noncontact method of measuring CCT and ACD. Larger differences occurred only when measuring peripheral CT values, especially between AC-Master and the other two methods.
比较三种不同测量角膜厚度(CT)和前房深度(ACD)的方法。
前瞻性临床试验(奥地利维也纳医科大学)。
使用Pentacam(旋转式Scheimpflug相机;德国韦茨拉尔Oculus公司)、Orbscan I(扫描裂隙地形图系统;美国犹他州盐湖城Orbtek公司)和AC-Master(部分相干干涉测量仪;德国耶拿蔡司医疗技术公司)对44名健康受试者的88只眼睛测量中央角膜厚度(CCT)、四个周边点的角膜厚度以及中央前房深度,并比较结果。
CCT测量值在AC-Master和Pentacam之间的一致性上限(下限)为7.9(-22.2)微米,在AC-Master和Orbscan之间为17.6(-32.5)微米,在Pentacam和Orbscan之间为25.2(-25.9)微米。三种方法之间的相关性都很高(r = 0.94至0.97)。ACD测量值在AC-Master和Pentacam之间的一致性上限和下限为0.174(-0.251)毫米,在AC-Master和Orbscan之间为0.406(-0.004)毫米,在Pentacam和Orbscan之间为0.384(0.095)毫米。三种方法之间的相关性很高(Orbscan和Pentacam之间r = 0.96;其他为0.92)。四个周边点的角膜厚度测量值的相关性较低。