Beyoglu Eye Research and Training Hospital, Istanbul, Turkey.
J Refract Surg. 2010 Feb;26(2):127-33. doi: 10.3928/1081597X-20100121-08. Epub 2010 Feb 12.
To evaluate the agreement among three different optical methods in measuring anterior chamber depth (ACD), central corneal thickness (CCT), and pupil diameter.
One hundred eyes of 50 healthy patients (25 men, 25 women) were enrolled in the study. Mean patient age was 25 years (range: 21 to 32 years). Exclusion criteria were history of any intraocular or corneal surgery, contact lens wear, corneal anomalies, and spherical refraction >5.00 diopters (D) or cylindrical refraction >2.00 D. All measurements were done by the same operator under mesopic light conditions and repeated using three different optical methods: Visante optical coherence tomography (OCT) (Carl Zeiss Meditec), Orbscan (Bausch & Lomb), and Pentacam (Oculus Optikgeräte GmbH).
Mean CCT as measured by Visante OCT, Orbscan, and Pentacam was 529+/-30.5 microm, 554+/-32.7 microm, and 552+/-29.3 microm, respectively. Mean ACD values were 2.94+/-0.34 mm, 2.84+/-0.33 mm, and 2.98+/-0.33 mm, respectively. Mean pupil diameter measurements were 4.87+/-1.09 mm, 4.0+/-0.67 mm, and 3.05+/-0.59 mm, respectively. The Visante OCT measured CCT thinner and Orbscan measured ACD shallower than the other two methods. All three methods measured pupil diameters significantly different.
This study found some statistically significant but clinically insignificant differences among the optical methods assessed. The differences are small and do not influence decisions for refractive surgery in clinical practice.
评估三种不同光学方法测量前房深度(ACD)、中央角膜厚度(CCT)和瞳孔直径的一致性。
本研究纳入了 50 名健康患者(25 名男性,25 名女性)的 100 只眼。患者平均年龄为 25 岁(范围:21 至 32 岁)。排除标准包括任何眼内或角膜手术史、隐形眼镜佩戴史、角膜异常以及球镜屈光度数>5.00 屈光度(D)或柱镜屈光度数>2.00 D。所有测量均由同一位操作者在中光条件下使用三种不同的光学方法进行:Visante 光学相干断层扫描(OCT)(卡尔蔡司医疗技术公司)、Orbscan(博士伦)和 Pentacam(欧科路光学设备有限公司)。
Visante OCT、Orbscan 和 Pentacam 测量的平均 CCT 分别为 529+/-30.5 微米、554+/-32.7 微米和 552+/-29.3 微米,平均 ACD 值分别为 2.94+/-0.34 毫米、2.84+/-0.33 毫米和 2.98+/-0.33 毫米,平均瞳孔直径测量值分别为 4.87+/-1.09 毫米、4.0+/-0.67 毫米和 3.05+/-0.59 毫米。Visante OCT 测量的 CCT 较薄,Orbscan 测量的 ACD 较浅。三种方法测量的瞳孔直径均有显著差异。
本研究发现评估的三种光学方法之间存在一些统计学上显著但临床上无意义的差异。这些差异较小,不会影响临床实践中屈光手术的决策。