Comparison of Anterior Segment Measurements with a New Multifunctional Unit and Five Other Devices.

机构信息

Department of Ophthalmology, Institute of Vision Research, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea Corneal Dystrophy Research Institute, Yonsei University College of Medicine, Seoul, Korea.

出版信息

Korean J Ophthalmol. 2022 Aug;36(4):338-349. doi: 10.3341/kjo.2022.0025. Epub 2022 Jun 15.

Abstract

PURPOSE

To evaluate the clinical availability of a multifunctional ocular biometric unit, MR-6000, for simultaneous keratometry, tonometry, topography, and pachymetry evaluation, and compare anterior segment measurements with five other devices: autokeratometer (KR-1), Scheimpflug camera (Pentacam HR), swept-source optical coherence tomography (IOLMaster 700), Placido disk scanning-slit topography (Orbscan II), and noncontact tonometry (FT-1000).

METHODS

Thirty eyes from thirty patients who visited Severance Hospital for cataract surgery were examined using MR6000 and the other devices. The mean keratometry, central corneal thickness (CCT), white-to-white (WTW) distance, and intraocular pressure (IOP) values were compared. Repeated measures analysis of variance, Wilcoxon signed-rank test, intraclass correlation coefficient (ICC), and Bland-Altman plot were used to assess the correlation and agreement between devices.

RESULTS

Thirty eyes of thirty patients were evaluated. Statistically significant differences in mean keratometry between MR6000, KR-1, Pentacam HR, and IOLMaster 700 were not observed (p > 0.05). All five devices, including Orbscan II, had almost perfect agreement in measuring keratometry (ICC > 0.80, p < 0.05). CCT measured by MR-6000 was significantly different from that of Pentacam HR and Orbscan II measurements (p < 0.05) but correlated with that of Pentacam HR and Orbscan II measurements (ICC > 0.60, p < 0.05). The WTW distance measured by MR-6000 was not significantly different from that measured by IOLMaster 700 but was different from that measured by Orbscan II. IOP measured by MR-6000 was not correlated with FT-1000.

CONCLUSIONS

Keratometric values obtained through MR-6000 can be used interchangeably with other devices based on good correlation and agreement. However, the CCT, WTW, and IOP values were not interchangeable with a single multifunctional unit for cataract surgery preoperative examination.

摘要

目的

评估一款多功能眼生物测量仪 MR-6000 的临床可用性,该仪器可同时进行角膜曲率、眼压、地形图和角膜厚度测量,并将眼前节测量值与其他五种设备进行比较:自动角膜曲率计(KR-1)、Scheimpflug 相机(Pentacam HR)、扫频源光学相干断层扫描仪(IOLMaster 700)、Placido 盘扫描裂隙地形图(Orbscan II)和非接触眼压计(FT-1000)。

方法

对 30 例在首尔峨山医院接受白内障手术的患者的 30 只眼进行了 MR6000 和其他设备的检查。比较平均角膜曲率、中央角膜厚度(CCT)、白对白(WTW)距离和眼压(IOP)值。采用重复测量方差分析、Wilcoxon 符号秩检验、组内相关系数(ICC)和 Bland-Altman 图评估设备之间的相关性和一致性。

结果

评估了 30 例 30 只眼。MR6000 与 KR-1、Pentacam HR 和 IOLMaster 700 之间的平均角膜曲率无统计学差异(p > 0.05)。所有五种设备(包括 Orbscan II)在测量角膜曲率方面具有几乎完美的一致性(ICC > 0.80,p < 0.05)。MR-6000 测量的 CCT 与 Pentacam HR 和 Orbscan II 测量值明显不同(p < 0.05),但与 Pentacam HR 和 Orbscan II 测量值相关(ICC > 0.60,p < 0.05)。MR-6000 测量的 WTW 距离与 IOLMaster 700 测量值无显著差异,但与 Orbscan II 测量值不同。MR-6000 测量的 IOP 与 FT-1000 不相关。

结论

基于良好的相关性和一致性,MR-6000 获得的角膜曲率值可与其他设备互换使用。然而,对于白内障术前检查,单个多功能单元无法互换使用 CCT、WTW 和 IOP 值。