Feldmann H J, Molls M, Hoederath A, Krümpelmann S, Sack H
Department of Radiooncology, West German Tumor Center, University Hospital, Essen.
Int J Radiat Oncol Biol Phys. 1992;23(5):1003-8. doi: 10.1016/0360-3016(92)90906-x.
Blood flow related data obtained in different deep-seated tumors and adjacent normal tissues were analyzed in 28 patients who were treated with combined regional hyperthermia and radiation for recurrent or metastatic tumors. The evaluation of blood flow related data has been made using the thermal clearance/thermal cooling coefficient technique and dynamic computed tomography. With both methods significant differences in global perfusion have been observed between tumor center and tumor periphery, between tumor and normal tissue (deep muscle and fat tissue), and between tumor entities. Washout rates or thermal cooling coefficient values, as well as the enhancement of contrast material over baseline (expressed in delta Hounsfield Units), correlated significantly with the achieved steady state temperatures for different tissue categories (i.e., tumor center, tumor periphery, different tumor entities, normal tissue). Thermal cooling coefficient values higher than 63000-83000 W/m3-K (washout rates higher than 15-20 ml/100 g-min) or values of enhancement of contrast material higher than delta 20-25 HU coincide with a limitation in achieving therapeutic temperatures higher than 40 degrees C.
对28例接受区域热疗联合放疗治疗复发性或转移性肿瘤的患者,分析了在不同深部肿瘤及相邻正常组织中获取的血流相关数据。已使用热清除/热冷却系数技术和动态计算机断层扫描对血流相关数据进行评估。通过这两种方法,观察到肿瘤中心与肿瘤周边、肿瘤与正常组织(深部肌肉和脂肪组织)以及不同肿瘤实体之间在整体灌注方面存在显著差异。洗脱率或热冷却系数值,以及对比剂相对于基线的增强(以亨氏单位变化表示),与不同组织类别(即肿瘤中心、肿瘤周边、不同肿瘤实体、正常组织)达到的稳态温度显著相关。热冷却系数值高于63000 - 83000 W/m³ - K(洗脱率高于15 - 20 ml/100 g - min)或对比剂增强值高于20 - 25 HU,与难以达到高于40摄氏度的治疗温度相关。