Tinkey Peggy, Lembo Thomas, Craig Suzanne, West Cheri, Van Pelt Carolyn
Department of Veterinary Medicine and Surgery, The University of Texas M.D. Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030, USA.
Lab Anim (NY). 2006 Feb;35(2):45-50. doi: 10.1038/laban0206-45.
The use of portable chemistry analyzers is an attractive option for obtaining clinical pathology panels in mice, because these analyzers require only small volumes of whole blood. However, in studies with other animals, portable analyzers do not always agree with results obtained using standard laboratory equipment. The authors evaluated the use of the i-STAT handheld portable clinical analyzer compared to the use of standard nonportable laboratory instruments in mice. As shown with other species, the i-STAT results did not always agree with standard laboratory instruments; however, the i-STAT does show reliability for certain chemistry assays.
对于从小鼠身上获取临床病理分析指标而言,使用便携式化学分析仪是一个颇具吸引力的选择,因为这些分析仪仅需少量全血。然而,在对其他动物的研究中,便携式分析仪的结果并不总是与使用标准实验室设备所获得的结果一致。作者评估了i-STAT手持式便携式临床分析仪与标准非便携式实验室仪器在小鼠研究中的使用情况。与其他物种的情况一样,i-STAT的结果并不总是与标准实验室仪器的结果一致;不过,i-STAT在某些化学分析中确实显示出可靠性。