Suppr超能文献

基于二维电容式微机械超声换能器(CMUT)阵列的三维声学成像前端集成电路设计。

Design of a front-end integrated circuit for 3D acoustic imaging using 2D CMUT arrays.

作者信息

Ciçek Ihsan, Bozkurt Ayhan, Karaman Mustafa

机构信息

Microelectronics Department, Faculty of Engineering, Sabanci University, Istanbul, Turkey.

出版信息

IEEE Trans Ultrason Ferroelectr Freq Control. 2005 Dec;52(12):2235-41. doi: 10.1109/tuffc.2005.1563266.

Abstract

Integration of front-end electronics with 2D capacitive micromachined ultrasonic transducer (CMUT) arrays has been a challenging issue due to the small element size and large channel count. We present design and verification of a front-end drive-readout integrated circuit for 3D ultrasonic imaging using 2D CMUT arrays. The circuit cell dedicated to a single CMUT array element consists of a high-voltage pulser and a low-noise readout amplifier. To analyze the circuit cell together with the CMUT element, we developed an electrical CMUT model with parameters derived through finite element analysis, and performed both the pre- and postlayout verification. An experimental chip consisting of 4 X 4 array of the designed circuit cells, each cell occupying a 200 X 200 microm2 area, was formed for the initial test studies and scheduled for fabrication in 0.8 microm, 50 V CMOS technology. The designed circuit is suitable for integration with CMUT arrays through flip-chip bonding and the CMUT-on-CMOS process.

摘要

由于元件尺寸小且通道数量多,将前端电子设备与二维电容式微机械超声换能器(CMUT)阵列集成一直是一个具有挑战性的问题。我们展示了一种使用二维CMUT阵列进行三维超声成像的前端驱动-读出集成电路的设计与验证。专用于单个CMUT阵列元件的电路单元由一个高压脉冲发生器和一个低噪声读出放大器组成。为了将电路单元与CMUT元件一起进行分析,我们开发了一个通过有限元分析得出参数的电CMUT模型,并进行了布局前和布局后的验证。为了进行初步测试研究,制作了一个由4×4阵列的设计电路单元组成的实验芯片,每个单元占据200×200平方微米的面积,并计划采用0.8微米、50V的CMOS技术进行制造。所设计的电路适合通过倒装芯片键合和CMOS上的CMUT工艺与CMUT阵列集成。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验