Crippa Michela, Belleri Luca, Mistrello Gianni, Tedoldi Chiara, Alessio Lorenzo
Institute of Occupational Health, University of Brescia, Regional hospital Sp. Civili of Brescia, Brescia, Italy.
Int Arch Occup Environ Health. 2006 Aug;79(7):550-7. doi: 10.1007/s00420-005-0080-5. Epub 2006 Feb 9.
In this study the latex protein content in devices commonly used in hospitals and general practice were investigated. The main aim was to acquire information for preventing latex allergy in health care workers and in the general population.
About 22 different types of medical devices and 23 devices commonly used in general practice were examined evaluating the total allergenic potency by a modified RAST-inhibition assay and quantitative determination of single allergens (Hev b1, Hev b5 and Hev b6.02) by using commercial ELISA kit.
A high level of inhibition was found in medical devices, such as elastic bandage (81.57%), tourniquet (74.09%), Foley urinary catheter (68.35%), Penrose drainage (67.25%) and taping (39.6%), and in common devices, such as rubber inner-sole (84.20%), toy balloon (78.62%), latex mattress (74.27%), household rubber gloves (49.10%), working gloves (38.25%), inflatable floating mattress (32.10%). Concentrations of latex extractable proteins and Hev b1, Hev b5 and Hev b6.02 antigens were high in some medical and general devices.
Latex exposure sources were found in hospitals and the home. These findings, though only preliminary and far from conclusive, could enable sensitized persons to avoid risky exposures and prevent allergic reactions. From the point of view of prevention, the time may come when every natural rubber object could be systematically labelled as "containing latex" together with the warning that "this item may cause allergic reactions in sensitized subjects."
在本研究中,对医院和普通诊所常用器械中的乳胶蛋白含量进行了调查。主要目的是获取信息,以预防医护人员和普通人群中的乳胶过敏。
检查了约22种不同类型的医疗设备和23种普通诊所常用的设备,通过改良的RAST抑制试验评估总变应原效力,并使用商用ELISA试剂盒对单一变应原(Hev b1、Hev b5和Hev b6.02)进行定量测定。
在医疗设备中,如弹性绷带(81.57%)、止血带(74.09%)、Foley导尿管(68.35%)、Penrose引流管(67.25%)和胶带(39.6%),以及普通设备中,如橡胶鞋垫(84.20%)、玩具气球(78.62%)、乳胶床垫(74.27%)、家用橡胶手套(49.10%)、工作手套(38.25%)、充气浮床(32.10%),发现了高水平的抑制。在一些医疗和普通设备中,可提取的乳胶蛋白以及Hev b1、Hev b5和Hev b6.02抗原的浓度很高。
在医院和家庭中都发现了乳胶暴露源。这些发现虽然只是初步的,远非结论性的,但可以使致敏者避免危险暴露并预防过敏反应。从预防的角度来看,可能会有这样一个时候,即每个天然橡胶制品都可以系统地标记为“含乳胶”,并附带警告“本物品可能会使致敏者产生过敏反应”。