Unal Orhan, Li Junwei, Cheng Weiguo, Yu Hyuk, Strother Charles M
Department of Medical Physics, University of Wisconsin, Madison, WI 53792-1590, USA.
J Magn Reson Imaging. 2006 May;23(5):763-9. doi: 10.1002/jmri.20555.
To investigate the potential utility of magnetic resonance (MR)-visible coatings for passive visualization of therapeutic endovascular devices such as catheters and guidewires.
Using a multistep coating process, gadolinium-based coatings were applied to commercially available off-the-shelf catheters and guidewires. These coated devices were imaged in phantoms made of fat-free yogurt, saline, and whole blood and also in live canine aorta on a 1.5-T cardiovascular MR scanner using T1-weighted two-dimensional radiofrequency (RF)-spoiled gradient-recalled echo, two-dimensional spin echo, and three-dimensional RF-spoiled gradient-recalled echo techniques.
Commercially available off-the shelf catheters (4, 5, and 6 French) and guidewires (0.038 inch) were clearly visualized in all phantoms and canine aorta and the coatings proved to be durable and imageable without degradation in signal intensity up to 24 hours. MR-visible coatings address some of the shortcomings that have previously limited the role of MR as a guidance tool.
Both in vitro and in vivo visualization of therapeutic endovascular devices coated with MR-visible coatings are found to be clinically viable.
研究磁共振(MR)可见涂层在对诸如导管和导丝等治疗性血管内装置进行被动可视化方面的潜在效用。
采用多步骤涂层工艺,将基于钆的涂层应用于市售现成的导管和导丝。使用T1加权二维射频(RF)扰相梯度回波、二维自旋回波和三维RF扰相梯度回波技术,在由无脂酸奶、盐水和全血制成的模型以及活体犬主动脉中,对这些涂覆装置进行成像,成像设备为1.5-T心血管MR扫描仪。
市售现成的导管(4、5和6法式)和导丝(0.038英寸)在所有模型和犬主动脉中均清晰可见,且涂层被证明具有耐久性,在长达24小时内信号强度无衰减且可成像。MR可见涂层解决了一些先前限制MR作为引导工具作用的缺点。
发现涂有MR可见涂层的治疗性血管内装置在体外和体内的可视化在临床上都是可行的。