Vacca Aldo
Royal Brisbane and Women's Hospital, Herston, Queensland, Australia.
Aust N Z J Obstet Gynaecol. 2006 Apr;46(2):124-7. doi: 10.1111/j.1479-828X.2006.00540.x.
Safety recommendations for vacuum-assisted delivery focus on limiting the number of pulls and the duration of the procedure. With the introduction of the Kiwi OmniCup vacuum extractor incorporating a Traction Force Indicator, it is now possible to reliably monitor and control the amount of traction force exerted on the fetal scalp. The OmniCup was used in a prospective observational study of 119 vacuum-assisted deliveries in nulliparous women to establish practical guidelines for safe practice in typical clinical settings.
真空辅助分娩的安全建议着重于限制牵引次数和操作持续时间。随着配备牵引力指示器的奇异果全能杯真空吸引器的推出,现在能够可靠地监测和控制施加在胎儿头皮上的牵引力大小。在一项针对119例初产妇真空辅助分娩的前瞻性观察研究中使用了全能杯,以建立典型临床环境中安全操作的实用指南。