The objective of our study was to investigate the impact of radial k-space sampling and steady-state free precession (SSFP) imaging on image quality in MRI of coronary vessel walls.
Eleven subjects were examined on a 1.5-T MR system using three high-resolution navigator-gated and cardiac-triggered 3D black blood sequences (cartesian gradient-echo [GRE], radial GRE, and radial SSFP) with identical spatial resolution (0.9 x 0.9 x 2.4 mm3). The signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), vessel wall sharpness, and motion artifacts were analyzed.
The mean SNR and CNR of the coronary vessel wall were improved using radial imaging and were best using radial k-space sampling combined with SSFP imaging. Vessel border definition was similar for all three sequences. Radial k-space sampling was found to be less sensitive to motion. Consistently good image quality was seen with the radial GRE sequence.
Radial k-space sampling in MRI of coronary vessel walls resulted in fewer motion artifacts and improved SNR and CNR. The use of SSFP imaging, however, did not result in improved coronary vessel wall visualization.
本研究的目的是探讨径向k空间采样和稳态自由进动(SSFP)成像对冠状动脉壁MRI图像质量的影响。
11名受试者在1.5-T MR系统上接受检查,使用三种具有相同空间分辨率(0.9×0.9×2.4 mm3)的高分辨率导航门控和心脏触发3D黑血序列(笛卡尔梯度回波[GRE]、径向GRE和径向SSFP)。分析信噪比(SNR)、对比噪声比(CNR)、血管壁清晰度和运动伪影。
使用径向成像可提高冠状动脉壁的平均SNR和CNR,采用径向k空间采样结合SSFP成像时效果最佳。所有三个序列的血管边界清晰度相似。发现径向k空间采样对运动不太敏感。径向GRE序列始终具有良好的图像质量。
冠状动脉壁MRI中的径向k空间采样减少了运动伪影,提高了SNR和CNR。然而,使用SSFP成像并未改善冠状动脉壁的可视化效果。