在使用AU5800生化分析仪对尿总蛋白检测进行携带污染评估时检测到累积IgG试剂携带污染。

Cumulative IgG reagent carryover detected in carryover evaluation of urinary total protein testing using AU5800 biochemistry analyzer.

作者信息

Li Jialu, Bao Guoxiang

机构信息

Clinical Laboratory Center, Shaoxing People's Hospital, Shaoxing, Zhejiang, China.

出版信息

Biochem Med (Zagreb). 2025 Jun 15;35(2):020702. doi: 10.11613/BM.2025.020702. Epub 2025 Apr 15.

Abstract

INTRODUCTION

Accurate measurement of urinary total protein (UTP) is crucial for diagnosing renal and systemic diseases. This study aims to comprehensively evaluate potential carryover scenarios in UTP testing using AU5800 biochemistry analyzer and to identify factors influencing assay accuracy.

MATERIALS AND METHODS

High-concentration quality control materials and pure water was used to evaluate specific sample probe carryover. Additionally, 24-hour mixed urine samples from patients were used to evaluate specific carryover from the reagent probe, mixbars, and cuvettes. For cumulative sample carryover evaluation, pure water was used as reagent and mixed serum as sample for continuous testing. During the process, 24-hour urine samples were interspersed, and UTP concentrations were measured at 0, 10, 20, and 30 minutes. Cumulative reagent carryover was evaluated by testing pure water sequentially with eight reagents sharing the same analytical unit and inner cuvettes as UTP, with UTP concentrations of 24-hour urine samples recorded at 0, 10, 20, and 30 minutes.

RESULTS

Specific carryover from the sample probe, reagent probe, mixbars, and cuvettes was not detected during carryover evaluation of UTP testing. However, a significant cumulative reagent carryover of Immunoglobulin G (IgG) reagent was observed, while no cumulative sample carryover was identified.

CONCLUSIONS

The full range of possible carryover scenarios in UTP testing was evaluated using AU5800 biochemistry analyzer. Our data provide valuable references for evaluating carryover in laboratories, ensuring the accuracy and reliability of laboratory results.

摘要

引言

准确测量尿总蛋白(UTP)对于诊断肾脏和全身性疾病至关重要。本研究旨在全面评估使用AU5800生化分析仪进行UTP检测时潜在的携带污染情况,并确定影响检测准确性的因素。

材料与方法

使用高浓度质量控制材料和纯水评估特定样本探针的携带污染情况。此外,还使用患者的24小时混合尿液样本评估试剂探针、搅拌棒和比色杯的特定携带污染情况。对于累积样本携带污染评估,使用纯水作为试剂,混合血清作为样本进行连续检测。在此过程中穿插加入24小时尿液样本,并在0、10、20和30分钟时测量UTP浓度。通过依次用八种与UTP共享同一分析单元和内部比色杯的试剂检测纯水来评估累积试剂携带污染情况,在0、10、20和30分钟时记录24小时尿液样本的UTP浓度。

结果

在UTP检测的携带污染评估过程中,未检测到样本探针、试剂探针、搅拌棒和比色杯的特定携带污染。然而,观察到免疫球蛋白G(IgG)试剂存在显著的累积试剂携带污染,而未发现累积样本携带污染。

结论

使用AU5800生化分析仪评估了UTP检测中所有可能的携带污染情况。我们的数据为实验室评估携带污染提供了有价值的参考,确保了实验室结果的准确性和可靠性。