Suppr超能文献

使用一种新的X射线技术测量髋臼杯的解剖学方位。

Measuring anatomical acetabular cup orientation with a new X-ray technique.

作者信息

Müller Otto, Reize Patrik, Trappmann Dirk, Wülker Nikolaus

机构信息

Orthopedic Clinic, University of Tübingen, Germany.

出版信息

Comput Aided Surg. 2006 Mar;11(2):69-75. doi: 10.3109/10929080600640618.

Abstract

Proper alignment of the acetabular cup component is one of the most important requisites for a successful long-term outcome in total hip replacement. However, measurement and indication of cup orientation in an anatomical pelvic reference system is very difficult. We propose a new C-arm-based X-ray technique for determining the values for inclination and anteversion of the acetabular cup component. The proposed method is validated by computer simulation and sources of error are evaluated. The method predicts an accuracy of better then 5 degrees for determination of anteversion of the cup.

摘要

髋臼杯组件的正确对线是全髋关节置换长期成功的最重要要求之一。然而,在解剖学骨盆参考系统中测量和指示髋臼杯的方向非常困难。我们提出了一种基于C型臂的新X射线技术来确定髋臼杯组件的倾斜度和前倾角值。通过计算机模拟对所提出的方法进行了验证,并评估了误差来源。该方法预测杯前倾角测定的准确度优于5度。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验