使用去白细胞血液进行左心室辅助装置植入后的致敏反应并非由HLA抗体引起。

Sensitization following LVAD implantation using leucodepleted blood is not due to HLA antibodies.

作者信息

Newell H, Smith J D, Rogers P, Birks E, Danskine A J, Fawson R E, Rose M L

机构信息

Transplant Immunology, Harefield Hospital, Harefield, Middx UB9 6JH, UK.

出版信息

Am J Transplant. 2006 Jul;6(7):1712-7. doi: 10.1111/j.1600-6143.2006.01342.x.

Abstract

Implantation of left ventricular assist devices (LVAD) is associated with HLA antibody sensitization. The objective of this study was to determine the specificity of antibodies produced by LVAD recipients using a combination of ELISA, Luminex and microcytotoxicity assays. Fifty-one LVAD patients were studied, from 44 to 838 days post-implantation. No patient developed HLA antibodies, although 24 produced IgG antibodies detectable in both ELISA and Luminex assays. These antibodies manifest as positive reactions with class I and class II wells of the ELISA and also blank wells. In Luminex assays, they produce high MFI readings with the negative control beads. Antibodies were detected 18 to 228 days after implantation. This reactivity was found to be directed against bovine serum albumin (BSA), commonly used to block non-specific binding in ELISA and Luminex assays; absorption of sera with BSA-coated beads completely abrogated reactivity in all solid phase assays, but did not eliminate anti-HLA antibodies in control sera. Ten of the 24 patients have proceeded to transplantation, with a 1-year graft survival of 69%. In conclusion, it appears that implantation of LVADS disrupts immunoregulatory pathways leading to production of anti-albumin antibodies. These can be misinterpreted as anti-HLA antibodies in solid phase assays.

摘要

植入左心室辅助装置(LVAD)与HLA抗体致敏相关。本研究的目的是使用酶联免疫吸附测定(ELISA)、Luminex检测和微细胞毒性测定相结合的方法,确定LVAD接受者产生的抗体的特异性。对51例LVAD患者进行了研究,这些患者在植入后44至838天。尽管有24例患者产生了在ELISA和Luminex检测中均可检测到的IgG抗体,但没有患者产生HLA抗体。这些抗体在ELISA的I类和II类孔以及空白孔中均表现为阳性反应。在Luminex检测中,它们与阴性对照微珠产生高平均荧光强度(MFI)读数。抗体在植入后18至228天被检测到。发现这种反应性针对牛血清白蛋白(BSA),BSA常用于阻断ELISA和Luminex检测中的非特异性结合;用包被BSA的微珠吸收血清可完全消除所有固相检测中的反应性,但不能消除对照血清中的抗HLA抗体。24例患者中有10例已接受移植,1年移植物存活率为69%。总之,似乎LVAD的植入破坏了免疫调节途径,导致抗白蛋白抗体的产生。在固相检测中,这些抗体可能会被误解为抗HLA抗体。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验