Qi Huili, Lv Jiagen, Li Baoxin
Department of Chemistry, School of Chemistry and Materials Science, Shaanxi Normal University, Xi'an 710062, PR China.
Spectrochim Acta A Mol Biomol Spectrosc. 2007 Apr;66(4-5):874-8. doi: 10.1016/j.saa.2006.05.003. Epub 2006 May 10.
In this paper, a simple flow-injection chemiluminescence (CL) system combined with on-line solid-phase extraction is presented to determine phenol. This method is based on the enhancement effect of phenol on the luminol-K3Fe(CN)6 CL system. The solid-phase extraction promised the high sensitivity and improved selectivity of CL detection. With the calibration range from 4.7 ng l-1 to 470 ng l-1 phenol concentration, the proposed method was applied to analyzing phenol in water samples and the obtained results were validated by the standard method. The detection limit was determined as 0.66 ng l-1. The relative standard deviation was 1.5% for determining 4.7 ng l-1 phenol standard (n=7).
本文提出了一种结合在线固相萃取的简单流动注射化学发光(CL)系统来测定苯酚。该方法基于苯酚对鲁米诺-K3Fe(CN)6化学发光系统的增强作用。固相萃取保证了化学发光检测的高灵敏度和选择性的提高。在苯酚浓度校准范围为4.7 ng l-1至470 ng l-1的情况下,该方法应用于分析水样中的苯酚,所得结果通过标准方法进行了验证。检测限确定为0.66 ng l-1。测定4.7 ng l-1苯酚标准品时的相对标准偏差为1.5%(n = 7)。