Bao P, Sinha T K, Chen C-C R, Warmath J R, Galloway R L, Herline A J
Vanderbilt University Medical Center, D-5220 Medical Center North, Nashville, TN 37232, USA.
Surg Endosc. 2007 Jan;21(1):74-9. doi: 10.1007/s00464-005-0220-8. Epub 2006 Oct 5.
Advanced laparoscopic procedures, particularly laparoscopic liver resection and ablation, may benefit from image-guided surgery techniques that involve interactive three-dimensional imaging and instrument tracking.
A prototype system for laparoscopic ultrasound-guided radiofrequency ablation was designed and implemented. This system uses an infrared camera to track instruments and runs on a personal computer. Features of the system include spatially registered ultrasound visualization, volume reconstruction, and interactive targeting. Targeting of accuracy studies was performed by directing a tracked needle to a phantom target.
Ultrasound data collection and volume reconstruction can be achieved within minutes and interactively reviewed by the surgeon. Early results with phantom experiments demonstrate a targeting accuracy of 5 to 10 mm.
These results support the further development of this and similar image-guided surgery systems for specific laparoscopic procedures. Eventually, rigorous clinical evaluation will be necessary to prove their value.