Ni Wendy, Christen Thomas, Zun Zungho, Zaharchuk Greg
Department of Radiology, Stanford University, Stanford, California, USA; Department of Electrical Engineering, Stanford University, Stanford, California, USA.
Magn Reson Med. 2015 Mar;73(3):1228-36. doi: 10.1002/mrm.25232. Epub 2014 Apr 18.
R2', the reversible component of transverse relaxation, is an important susceptibility measurement for studies of brain physiology and pathologies. In existing literature, different R2' measurement methods are used with assumption of equivalency. This study explores the choice of measurement method in healthy, young subjects at 3T.
In this study, a modified gradient-echo sampling of free induction decay and echo (GESFIDE) sequence was used to compare four standard R2' measurement methods: asymmetric spin echo (ASE), standard GESFIDE, gradient echo sampling of the spin echo (GESSE), and separate R2 and R2* mapping.
GESSE returned lower R2' measurements than other methods (P < 0.05). Intersubject mean R2' in gray matter was found to be 2.7 s(-1) using standard GESFIDE and GESSE, versus 3.4-3.8 s(-1) using other methods. In white matter, mean R2' from GESSE was 2.3 s(-1) while other methods produced 3.7-4.3 s(-1) . R2 correction was applied to partially reduce the discrepancies between the methods, but significant differences remained, likely due to violation of the fundamental assumption of a single-compartmental tissue model, and hence monoexponential decay.
R2' measurements are influenced significantly by the choice of method. Awareness of this issue is important when designing and interpreting studies that involve R2' measurements.
横向弛豫的可逆成分R2′是研究脑生理学和病理学的一项重要的磁化率测量指标。在现有文献中,不同的R2′测量方法在假定等效性的情况下被使用。本研究探讨了在3T场强下健康年轻受试者中测量方法的选择。
在本研究中,采用一种改进的自由感应衰减和回波梯度回波采样(GESFIDE)序列来比较四种标准的R2′测量方法:非对称自旋回波(ASE)、标准GESFIDE、自旋回波梯度回波采样(GESSE)以及单独的R2和R2*映射。
GESSE得出的R2′测量值低于其他方法(P<0.05)。使用标准GESFIDE和GESSE时,灰质的受试者间平均R2′为2.7 s⁻¹,而使用其他方法时为3.4 - 3.8 s⁻¹。在白质中,GESSE得出的平均R2′为2.3 s⁻¹,而其他方法得出的为3.7 - 4.3 s⁻¹。应用R2校正以部分减少方法之间的差异,但仍存在显著差异,这可能是由于违反了单室组织模型的基本假设,进而违反了单指数衰减的假设。
R2′测量值受测量方法选择的显著影响。在设计和解释涉及R2′测量的研究时,意识到这个问题很重要。