Titanium alloy mini-implants for orthodontic anchorage: immediate loading and metal ion release.

作者信息

Morais Liliane S, Serra Glaucio G, Muller Carlos A, Andrade Leonardo R, Palermo Elisabete F A, Elias Carlos N, Meyers Marc

机构信息

Mechanical and Aerospace Engineering Department, University of California-San Diego (UCSD), San Diego, CA, United States.

出版信息

Acta Biomater. 2007 May;3(3):331-9. doi: 10.1016/j.actbio.2006.10.010. Epub 2007 Jan 25.

Abstract

Removable osseointegrated titanium mini-implants were successfully used as anchorage devices in orthodontics. The early load is necessary to simplify the mini-implant methodology, but can lead to failure during osseointegration. The Ti-6Al-4V alloy was used instead of commercially pure Ti due to its superior strength. However, the corrosion resistance is low, allowing for metal ion release. The purpose of this work was to analyze the immediately loaded mini-implant fixation and to gauge the vanadium ion release during the healing process. Titanium alloy mini-implants were inserted in the tibiae of rabbits. After 1, 4 and 12 weeks, they were submitted to removal torque testing. There was no increase in the removal torque value between 1 and 4 weeks of healing, regardless of the load. Nevertheless, after 12 weeks, a significant improvement was observed in both groups, with the highest removal torque value for the unloaded group. The kidney, liver and lung were also extracted and analyzed by atomic absorption spectrometry. In comparison with the control values, the content of vanadium increased slightly after 1week, significantly increased after 4 weeks and decreased slightly after 12 weeks, without reaching the 1 week values. A stress analysis was carried out which enables both the prediction of the torque at which commercially pure (CP) Ti and Ti-6Al-4V deform plastically and the shear strength of the interface. This analysis reveals that the removal torques for CP Ti dangerously approach the yield stress. The results of this rabbit model study indicate that titanium alloy mini-implants can be loaded immediately with no compromise in their stability. The detected concentration of vanadium did not reach toxic levels in the animal model.

摘要

可摘式骨结合钛微型种植体已成功用作正畸治疗中的支抗装置。早期加载对于简化微型种植体方法是必要的,但可能导致骨结合过程中的失败。由于其优越的强度,Ti-6Al-4V合金被用于替代商业纯钛。然而,其耐腐蚀性较低,会导致金属离子释放。这项工作的目的是分析即刻加载的微型种植体固定情况,并测量愈合过程中钒离子的释放。将钛合金微型种植体植入兔胫骨。在1、4和12周后,对其进行去除扭矩测试。在愈合的1至4周内,无论是否加载,去除扭矩值均未增加。然而,12周后,两组均观察到显著改善,未加载组的去除扭矩值最高。还提取了肾脏、肝脏和肺,并通过原子吸收光谱法进行分析。与对照值相比,钒含量在1周后略有增加,4周后显著增加,12周后略有下降,但未达到1周时的值。进行了应力分析,该分析能够预测商业纯钛(CP Ti)和Ti-6Al-4V发生塑性变形时的扭矩以及界面的剪切强度。该分析表明,CP Ti的去除扭矩危险地接近屈服应力。该兔模型研究结果表明,钛合金微型种植体可以即刻加载,而不会影响其稳定性。在动物模型中检测到的钒浓度未达到有毒水平。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验