犬股骨皮质骨中正畸钛钉微型种植体初始稳定性的扭矩去除研究

A torque removal study on the primary stability of orthodontic titanium screw mini-implants in the cortical bone of dog femurs.

作者信息

Okazaki J, Komasa Y, Sakai D, Kamada A, Ikeo T, Toda I, Suwa F, Inoue M, Etoh T

机构信息

Department of Geriatric Dentistry, Osaka Dental University, Japan.

出版信息

Int J Oral Maxillofac Surg. 2008 Jul;37(7):647-50. doi: 10.1016/j.ijom.2008.04.007. Epub 2008 Jun 12.

Abstract

The aim of this study was to biomechanically evaluate the primary stability of pure titanium orthodontic mini-implants, inserted into pre-drilled cavities of differing diameters. Mini-implants (1.2 mm diameter) were placed into 1.0 mm and 1.2 mm diameter cavities prepared in the mid-region of the bilateral hind leg femurs of anesthetized beagles. Removal torque strengths were measured immediately, 1, 3, 6, 9 and 12 weeks post-insertion of the implant. For mini-implants placed into 1-mm cavities, removal torque values decrease over the first 6 weeks (p<0.01), after which values remained static. Average values obtained immediately, 1, 3 and 6 weeks post-insertion were 10.98, 8.83, 7.20 and 5.12 Ncm, respectively . Immediately post-insertion, removal torque values of mini-implants placed in a 1.2-mm cavity, were 11-fold lower than those placed in 1.0-mm cavities, which then demonstrated a significant increase in strength from 3 weeks (1.35 Ncm) to 6 weeks (5.17 Ncm) post-insertion (p<0.01). Measurements 6, 9 and 12 weeks post-insertion were similar to those in the 1.0-mm cavity. Initial stability of titanium mini-implants is considered necessary for immediate and early use in orthodontics, and an implant without this initial stability should be replaced or isolated until it develops the appropriate stability supported by osseointegration.

摘要

本研究的目的是对插入不同直径预钻孔洞的纯钛正畸微型种植体的初期稳定性进行生物力学评估。将微型种植体(直径1.2毫米)植入麻醉后的比格犬双侧后腿股骨中部制备的直径1.0毫米和1.2毫米的孔洞中。在植入种植体后立即、1、3、6、9和12周测量去除扭矩强度。对于植入1毫米孔洞中的微型种植体,去除扭矩值在最初6周内下降(p<0.01),此后数值保持稳定。植入后立即、1、3和6周获得的平均值分别为10.98、8.83、7.20和5.12牛厘米。植入后立即,植入1.2毫米孔洞中的微型种植体的去除扭矩值比植入1.0毫米孔洞中的低11倍,然后从植入后3周(1.35牛厘米)到6周(5.17牛厘米)强度显著增加(p<0.01)。植入后6、9和12周的测量结果与1.0毫米孔洞中的相似。钛微型种植体的初期稳定性被认为是在正畸中立即和早期使用所必需的,并且没有这种初期稳定性的种植体应该被替换或隔离,直到它通过骨整合获得适当的稳定性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验