Validation of 133Xe clearance as a cerebral blood flow measurement technique during cardiopulmonary bypass.

作者信息

Spahn D R, Quill T J, Hu W C, Lu J, Smith L R, Reves J G, McRae R L, Leone B J

机构信息

Department of Anesthesiology, Duke University Medical Center, Durham, North Carolina 27710.

出版信息

J Cereb Blood Flow Metab. 1992 Jan;12(1):155-61. doi: 10.1038/jcbfm.1992.19.

Abstract

133Xe clearance to measure cerebral blood flow (CBF) was examined in 10 dogs during cardiopulmonary bypass. As a reference method, a continuous Kety-Schmidt technique (CBFKS) with 133Xe as indicator was used. Extracranial tissue was removed to directly place the 133Xe detectors on the skull, and the head was covered with a 3 mm lead shield to minimize contamination of the 133Xe clearance curve with extracranial radiation. 133Xe detectors for the Kety-Schmidt technique were embedded in a shielded brass block to minimize interference with radiation from the animal's body. 133Xe clearance data were analyzed using stochastic (CBF10, CBF15, and CBFINF) and initial slope methods (CBFIS), and the results were compared with CBFKS using linear regression. CBF15 and CBFINF yielded similar CBF values as CBFKS (CBFKS = 0.97.CBF15-2.08, r = 0.92, p less than 0.01; CBFKS = 1.13.CBFINF-1.21, r = 0.92, p less than 0.01). CBF10 slightly overestimated CBFKS but still showed a close correlation to CBFKS (CBFKS = 0.89.CBF10-2.58, r = 0.92, p less than 0.01) and CBFIS considerably overestimated CBFKS (CBFKS = 0.60.CBFIS-1.27, r = 0.87, p less than 0.01). With extracranial contamination of the 133Xe clearance curve minimized, all 133Xe clearance techniques used to measure CBF were consistently related to CBFKS in a constant, significant manner. 133Xe clearance therefore is a valid method to assess CBF during cardiopulmonary bypass.

摘要

相似文献

1
Validation of 133Xe clearance as a cerebral blood flow measurement technique during cardiopulmonary bypass.
J Cereb Blood Flow Metab. 1992 Jan;12(1):155-61. doi: 10.1038/jcbfm.1992.19.
6
The effect of hematocrit and systolic blood pressure on cerebral blood flow in newborn infants.
J Cereb Blood Flow Metab. 1987 Jun;7(3):295-9. doi: 10.1038/jcbfm.1987.66.
8
Monoexponential analysis of 133Xe clearance curves for regional cerebral blood flow measurements.
J Cereb Blood Flow Metab. 1984 Jun;4(2):250-8. doi: 10.1038/jcbfm.1984.34.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验