Song Shujuan, Song Qi Jun, Chen Zhongliang
School of Chemical and Material Engineering, Southern Yangtze University, Wuxi 214122, China.
Anal Bioanal Chem. 2007 Apr;387(8):2917-22. doi: 10.1007/s00216-007-1130-5. Epub 2007 Feb 9.
A highly selective and sensitive chemiluminescence method for the determination of triclosan is proposed. The method is based on the phototransformation of triclosan to a light-emitting precursor in the presence of fluorescein in alkaline medium and the chemiluminescence reaction is then triggered by strong base or oxidants such as N-bromosuccinimide. Based on this reaction an online phototransformation-flow injection manifold was developed, in which the photoreactor comprises a 150-cm-long x 0.8-mm-i.d. piece of PTFE tubing coiled around a 25-W fluorescent lamp, and the phototransformed products were then injected into a carrier stream of borate buffer. After mixing with the oxidant stream the produced light was detected by a photomultiplier. A wide calibration range from 8.0 x 10(-8) to 1.0 x 10(-4) mol L(-1) was obtained under the optimized conditions, and the detection limit was as low as 5.0 x 10(-8) mol L(-1). The whole process of analysis, including the online phototransformation and subsequent chemiluminescence detection, could be completed in 6 min. Most of the foreign substances tested showed high tolerance levels, and the proposed method was directly applied to the determination of triclosan in toothpaste samples without any pre-separation procedure. Figure Schematic representation of the phototransformation of triclosan and subsequent chemiluminescence reaction.