Clarke L P, Cullom S J, Shaw R, Reece C, Penney B C, King M A, Silbiger M
Department of Radiology, College of Medicine, University of South Florida, Tampa 33612.
J Nucl Med. 1992 Jan;33(1):161-6.
Quantitative imaging of bremsstrahlung from pure beta emitters is proposed as a means for in vivo management of antibody therapy. The method involves the use of high-energy collimation, an empirically selected broad photon energy window to enhance detector sensitivity, and a Wiener restoration filter to compensate for system blur. The measured and filtered data were obtained for an idealized scattering medium and isolated spherical sources. An effective linear attenuation coefficient of about 0.13 cm-1 was determined from the raw image data of 32P. A coefficient of 0.14 cm-1 was determined after the images were restored using the Wiener filter. The measured attenuation was not significantly dependent on the size of the region of interest or the size of the source. Its variation was within the experimental error of measurement (+/- 5%). The measured sensitivity (6 x 10(-6) cps/Bq) was sufficient for imaging therapy doses of 32P or 90Y.
提出将纯β发射体的轫致辐射定量成像作为体内抗体治疗管理的一种手段。该方法包括使用高能准直、经验选择的宽光子能量窗口以提高探测器灵敏度,以及使用维纳恢复滤波器来补偿系统模糊。针对理想化散射介质和孤立球形源获取了测量和滤波后的数据。从32P的原始图像数据确定有效线性衰减系数约为0.13 cm-1。使用维纳滤波器恢复图像后确定的系数为0.14 cm-1。测量的衰减对感兴趣区域的大小或源的大小没有显著依赖性。其变化在测量的实验误差(±5%)范围内。测量的灵敏度(6×10(-6) cps/Bq)足以对32P或90Y的治疗剂量进行成像。