一种用于在左心房-肺静脉交界处进行标测和射频消融的新型网状电极导管:一种单导管肺静脉前庭隔离方法。

A novel mesh electrode catheter for mapping and radiofrequency delivery at the left atrium-pulmonary vein junction: a single-catheter approach to pulmonary vein antrum isolation.

作者信息

Arruda Mauricio S, He Ding Sheng, Friedman Paul, Nakagawa Hiroshi, Bruce Charles, Azegami Koji, Anders Robert, Kozel Peter, Chiavetta Amedeo, Marad Paul, MacAdam David, Jackman Warren, Wilber David J

机构信息

Cleveland Clinic Foundation, Cleveland, Ohio 44195, USA.

出版信息

J Cardiovasc Electrophysiol. 2007 Feb;18(2):206-11. doi: 10.1111/j.1540-8167.2007.00720.x.

Abstract

BACKGROUND

Electrical isolation of pulmonary veins (PV) by radiofrequency (RF) ablation is often performed in patients with atrial fibrillation (AF). Current catheter technology usually requires the use of a multielectrode catheter for mapping in addition to the ablation catheter.

PURPOSE

We evaluated the feasibility and safety of using a single, expandable electrode catheter (MESH) to map and to electrically isolate the PV.

METHODS AND RESULTS

Nineteen closed-chest mongrel dogs, weighing 23-35 kg, were studied under general anesthesia. Intracardiac echocardiography (ICE) was used to guide transseptal puncture and to assess PV dimensions and contact of the MESH with PV ostia. ICE and angiography of RSPV were obtained before and after ablation, and prior to sacrifice at 7-99 days. An 11.5 Fr steerable MESH was advanced and deployed at the ostium of the RSPV. Recordings were obtained via the 36 electrodes comprising the MESH. For circumferential ablation, RF current was delivered at a target temperature of 62-65 degrees C (4 thermocouples) and maximum power of 70-100 W for 180 to 300 seconds. Each animal received 1-4 RF applications. Entrance conduction block was obtained in 13/19 treated RSPVs. Pathological examination confirmed circumferential and transmural lesions in 13 of 19 RSPV. LA mural thrombus was present in 3 animals. There was no significant PV stenosis.

CONCLUSION

Based on this canine model, a new expandable MESH catheter may safely be used for mapping and for PV antrum isolation. This approach may decrease procedure time without compromising success rate in patients undergoing AF ablation.

摘要

背景

心房颤动(AF)患者常采用射频(RF)消融术进行肺静脉(PV)电隔离。目前的导管技术通常除了消融导管外,还需要使用多电极导管进行标测。

目的

我们评估了使用单一可扩张电极导管(MESH)进行肺静脉标测和电隔离的可行性和安全性。

方法与结果

对19只体重23 - 35千克的闭胸杂种犬在全身麻醉下进行研究。心内超声心动图(ICE)用于引导经房间隔穿刺,并评估肺静脉尺寸以及MESH与肺静脉口的接触情况。在消融前后以及7 - 99天处死前获取右上肺静脉(RSPV)的ICE和血管造影图像。将一个11.5 Fr的可操控MESH推进并部署在RSPV口。通过构成MESH的36个电极进行记录。对于环周消融,在目标温度62 - 65摄氏度(4个热电偶)和最大功率70 - 100瓦下施加射频电流180至300秒。每只动物接受1 - 4次射频应用。19只接受治疗的RSPV中有13只实现了入口传导阻滞。病理检查证实19只RSPV中有13只存在环周和透壁性病变。3只动物存在左心房壁血栓。未出现明显的肺静脉狭窄。

结论

基于此犬类模型,一种新型可扩张MESH导管可安全地用于标测和肺静脉前庭隔离。这种方法可能会减少手术时间,而不会降低接受房颤消融术患者的成功率。