Bal B Sonny, Greenberg David
Department of Orthopaedic Surgery, University of Missouri-Columbia, Columbia, Missouri, USA.
J Arthroplasty. 2007 Apr;22(3):464-7. doi: 10.1016/j.arth.2006.04.011.
Failures of the polyethylene tibial post in posterior stabilized total knee arthroplasty are usually associated with pain and knee instability. We report an unusual presentation of a tibial post failure that occurred after the polyethylene insert was revised in a posterior-stabilized total knee. The tibial post on the revised insert broke off, exposing a metal reinforcing pin. During the resulting posterior subluxations of the tibia, the metal pin articulated against the cam on the femoral component, resulting in the generation of metallic debris in the knee joint.
后稳定型全膝关节置换术中聚乙烯胫骨柱的失败通常与疼痛和膝关节不稳定有关。我们报告了一例不寻常的胫骨柱失败病例,该病例发生在对后稳定型全膝关节置换术的聚乙烯衬垫进行翻修之后。翻修后的衬垫上的胫骨柱折断,露出金属加强销。在由此导致的胫骨后脱位过程中,金属销与股骨部件上的凸轮相互铰接,导致膝关节内产生金属碎屑。