Suppr超能文献

一名9岁患者后稳定型膝关节假体中聚乙烯柱的两阶段骨折:病例报告

Two stage fracture of a polyethylene post in a 9-year-old posterior-stabilized knee prosthesis: a case report.

作者信息

D'Angelo Fabio, Marcolli Daniele, Bulgheroni Paolo, Murena Luigi, Congiu Terenzio, Cherubino Paolo

机构信息

Department of Orthopaedics and Traumatology, University of Insubria, Ospedale di Circolo - Fondazione Macchi, V le Borri 57, 21100 Varese, Italy.

出版信息

J Med Case Rep. 2010 Feb 23;4:65. doi: 10.1186/1752-1947-4-65.

Abstract

INTRODUCTION

Several cases of tibial post breakage are reported in the literature. To the best of our knowledge, only three cases of NexGen knee prosthesis (Zimmer, Warsaw, Indiana, USA) tibial post failure have been reported.

CASE PRESENTATION

In November 1999, a 63-year-old Caucasian woman from Italy with a history of symptomatic left knee osteoarthritis underwent a total knee arthroplasty. In March 2008, while rising from a chair, she felt a sudden pain and instability in her left knee. She reported a fracture of the polyethylene post of the tibial insert. No malposition or malalignment of either the femoral or tibial components were identified. The polyethylene tibial insert was studied under light microscopy and scanning electron microscopy. The fracture was also noted to have occurred without any notable polyethylene wear.

CONCLUSION

Scanning electron microscopy revealed two different damage patterns that could be explained with a two-stage rupture of our patient's polyethylene post. This could have been caused by a non-optimal ligamentous balancing during first implant surgery. Her knee probably developed a varus instability that weakened the post, and then a posterior anterior stress finally broke the polyethylene.

摘要

引言

文献中报道了几例胫骨柱断裂的病例。据我们所知,仅报道了3例NexGen膝关节假体(美国印第安纳州华沙市的捷迈公司生产)胫骨柱失效的病例。

病例介绍

1999年11月,一名来自意大利的63岁白人女性,有症状性左膝骨关节炎病史,接受了全膝关节置换术。2008年3月,当她从椅子上起身时,感到左膝突然疼痛且不稳定。她报告胫骨植入物的聚乙烯柱发生了骨折。未发现股骨或胫骨部件有位置不当或排列不齐的情况。对聚乙烯胫骨植入物进行了光学显微镜和扫描电子显微镜检查。还注意到骨折发生时没有任何明显的聚乙烯磨损。

结论

扫描电子显微镜显示了两种不同的损伤模式,这可以用我们患者聚乙烯柱的两阶段破裂来解释。这可能是由于初次植入手术时韧带平衡不理想所致。她的膝盖可能出现了内翻不稳定,削弱了柱体,然后前后应力最终导致聚乙烯断裂。