经腹段下腔静脉测量中心静脉压的准确性。

Accuracy of central venous pressure measurement from the abdominal inferior vena cava.

作者信息

Lloyd T R, Donnerstein R L, Berg R A

机构信息

University Heart Center, Steele Memorial Children's Research Center, University of Arizona College of Medicine, Tucson.

出版信息

Pediatrics. 1992 Mar;89(3):506-8.

PMID:1741229
Abstract

Central venous pressure measurements in the abdominal inferior vena cava were compared with measurements in the right atrium in 10 infants and 10 children during cardiac catheterization. At end expiration, the mean pressures at these two sites were within 1 mm Hg of each other in all 20 patients, with a mean difference of 0.0 +/- 0.36 mm Hg. The abdominal inferior vena cava is a safe and convenient site for measurement of central venous pressure, and our study confirms that such measurements are accurate.

摘要

在10名婴儿和10名儿童进行心导管插入术期间,对腹部下腔静脉的中心静脉压测量值与右心房的测量值进行了比较。在呼气末,所有20例患者这两个部位的平均压力相差不超过1毫米汞柱,平均差值为0.0±0.36毫米汞柱。腹部下腔静脉是测量中心静脉压的一个安全且方便的部位,我们的研究证实了这种测量是准确的。