Schulz-Wendtland R, Hermann K-P, Wenkel E, Adamietz B, Lell M, Anders K, Uder M
Abteilung Gynäkologische Radiologie, Radiologisches Institut, Universitätsklinikum Erlangen, Universitätsstr. 21-23, 91056 Erlangen, Deutschland.
Radiologe. 2011 Feb;51(2):130-4. doi: 10.1007/s00117-010-2077-7.
The article describes an experimental phantom study of a system for digital full field mammography with a new digital detector with a double plate of pure selenium.
The experiments were carried out with the new full field digital mammography system Amulet from FujiFilm. This system has a new detector (18×24 cm(2)) on the basis of highly purified amorphous selenium (a-Se) with a pixel size of 50 µm. The x-rays are converted into electric signals in the first plate which are read into the second plate with the help of an optical switch and demonstrated in the form of an image. In this way a better pixel size/volume and signal-to-noise ratio should be achieved. The object of the investigation was the Wisconsin Mammographic Random Phantom, Model 152 A (Technical Performance Mo/Mo, 28 kV, 100 mAs). Five investigators with different experiences in mammography each received three images on a monitor with different random positions of the simulated lesions in the phantom for assessment. The detection rates were compared under the same conditions with the results of two other full field digital mammography systems.
The median detection rate for all images and investigators for the new doubled plated a-Se detector with optical switch was 98.7%. For both other systems with a-Si or and a-Se detectors the detection rate was 89.8% or 97.3%, respectively. There were no significant differences in the detection rate of the simulated breast lesions for all three systems considering the interobserver and intraobserver variation.
The first phantom study for the detection of simulated breast lesions with the new full field digital mammography system Amulet demonstrates equivalent results with the other systems used in the clinical routine. The trend towards superiority of the new system has to be confirmed in further clinical studies.
本文描述了一种使用新型双纯硒板数字探测器的数字全场乳腺摄影系统的实验体模研究。
实验采用富士胶片公司的新型全场数字乳腺摄影系统Amulet进行。该系统有一个基于高纯度非晶硒(a-Se)的新型探测器(18×24 cm²),像素尺寸为50 µm。X射线在第一块板中转换为电信号,借助光开关读入第二块板,并以图像形式显示。通过这种方式应能实现更好的像素尺寸/体积和信噪比。研究对象是威斯康星乳腺摄影随机体模,型号152 A(技术性能Mo/Mo,28 kV,100 mAs)。五名在乳腺摄影方面经验不同的研究者,每人在监视器上接收三张图像,体模中模拟病变处于不同随机位置以供评估。在相同条件下,将检测率与另外两种全场数字乳腺摄影系统的结果进行比较。
对于配备光开关的新型双板a-Se探测器,所有图像和研究者的中位检测率为98.7%。对于另外两种分别使用非晶硅(a-Si)或a-Se探测器的系统,检测率分别为89.8%和97.3%。考虑到观察者间和观察者内的差异,对于所有三个系统,模拟乳腺病变的检测率没有显著差异。
首次使用新型全场数字乳腺摄影系统Amulet检测模拟乳腺病变的体模研究表明,其结果与临床常规使用的其他系统相当。新系统的优势趋势有待进一步临床研究证实。