Skirting: a limitation for the performance of X-ray microanalysis in the variable pressure or environmental scanning electron microscope.

作者信息

Le Berre J F, Demopoulos G P, Gauvin R

机构信息

Department of Mining, Metals and Materials Engineering, McGill University, M H Wong Building, Montreal, QC, Canada.

出版信息

Scanning. 2007 May-Jun;29(3):114-22. doi: 10.1002/sca.20052.

Abstract

The variable pressure or environmental scanning electron microscope (VP-SEM; ESEM) has become the microscope of choice for many scientists and technologists. Hence, the development of robust methods for X-ray microanalysis, limited by skirting, has become critical. In this paper, two pressure variation correction methods (Doehne and Gauvin) are compared. Both of these methods appear to be effective; the results were found to be well within 10% of the values obtained at 0 Pa. The Doehne method is dependent on an empirical factor (D), therefore the accuracy of the results will depend on the accuracy of this value. Also the Doehne method is compromised by the nonlinearity of the response with pressure. The Gauvin method is more user-friendly and more precise when considering the total range of pressure.