Miniature membrane oxygenator for use in neonatal ECMO.

作者信息

Katayama M, Miyasaka K, Hirose I, Shimada M, Mimura R, Akasu H

机构信息

Pathophysiology Research Laboratory, National Children's Medical Center, Tokyo, Japan.

出版信息

ASAIO Trans. 1991 Jul-Sep;37(3):M407-8.

PMID:1751211
Abstract

A new miniature membrane oxygenator (Kuraray KMO, size, 0.3 m2, with a priming volume of 47 ml, compliance of less than 0.1 ml/100 mmHg, and pressure loss of 45 mmHg) with improved gas transfer and mechanical durability was developed and tested. The membrane material is a hollow fiber double layer polyolefin. The testing procedures determined by the AAMI were followed, and the results showed improved O2 and CO2 transfer (70 ml/min and 55 ml/min, respectively). Hemolysis was within acceptable limits, and plasma leakage was undetectable after 7 days of perfusion. Clinical study demonstrated satisfactory performance.

摘要

一种新型微型膜式氧合器(可乐丽KMO,尺寸为0.3平方米,预充量为47毫升,顺应性小于0.1毫升/100毫米汞柱,压力损失为45毫米汞柱)被研发并测试,其气体传输和机械耐久性得到了改善。膜材料为中空纤维双层聚烯烃。遵循了美国医疗器械促进协会(AAMI)确定的测试程序,结果显示氧气和二氧化碳传输得到改善(分别为70毫升/分钟和55毫升/分钟)。溶血在可接受范围内,灌注7天后未检测到血浆渗漏。临床研究表明其性能令人满意。