Suppr超能文献

首日校正对系统设置误差减少的影响。

Effect of the first day correction on systematic setup error reduction.

作者信息

Wu Qiuwen, Lockman David, Wong John, Yan Di

机构信息

Department of Radiation Oncology, William Beaumont Hospital, Royal Oak, Michigan 48073, USA.

出版信息

Med Phys. 2007 May;34(5):1789-96. doi: 10.1118/1.2727299.

Abstract

Treatment simulation is usually performed with a conventional simulator using kV X-rays or with a computed tomography (CT) simulator before the treatment course begins. The purpose is to verify patient setup under the same conditions as for treatment planning. Systematic (preparation) setup errors can be introduced by this process. The purpose of this study is to characterize the setup errors using electronic portal image (EPI) analyses and to propose a method to reduce the systematic component by performing simulation and patient preparation on the treatment machine. In this study, the first four or five days EPIs were analyzed from a total of 533 prostate cancer patients who were simulated on conventional simulators. We characterized setup errors using four parameters: {M(microi), Sigma (microi), RMS(microi), sigma (sigmai)}, where microi and sigmai are individual patient mean and standard deviation, M, Sigma, and RMS are the mean, standard deviation, and root-mean-square of underlying variables (microi and sigmai). We have performed a simulation of removing systematic components by correcting the first day setup error. As a comparison, we also carried out a similar analyses for patients simulated on a CT simulator and patients treated on a linac with an on-board kV CT imaging system, although a limited number of patients were available in these two samples. We found that Sigma (/ui)=(2.6,3.4,2.4) mm, and RMS(sigmai)=(1.5,1.9,1.0) mm in lateral, anterior/posterior, and cranial/caudal directions, indicating that systematic errors are much larger than random errors. Strong correlations were found between measurement on the first day and microi, implying the first day's measurement is a good predictor for microi. The same parameters were also computed for days 2-4, with and without the first day correction. Without correction, M(microi)2-4=(0.7,1.6,-1.0) mm, and Sigma(microi)2-4=(2.6,3.5,2.4) mm. With correction, M(microi)2-4=(0.0,0.4,0.4) mm, much closer to zero, and Sigma(microi)2-4=(1.8,2.2, 1.2) mm, also much smaller. While the use of a CT simulator can reduce the systematic errors, the benefits of first day correction can still be observed, although at a smaller magnitude. Therefore, the systematic setup error can be significantly reduced if the patient is marked and fields are verified on the treatment machine on the first fraction, preferably with an on-board kV imaging system.

摘要

在治疗疗程开始前,通常使用传统模拟器(利用千伏级X射线)或计算机断层扫描(CT)模拟器进行治疗模拟。目的是在与治疗计划相同的条件下验证患者的摆位。此过程可能会引入系统性(准备阶段)摆位误差。本研究的目的是通过电子射野影像(EPI)分析来表征摆位误差,并提出一种通过在治疗机上进行模拟和患者准备来减少系统性误差成分的方法。在本研究中,分析了总共533例在传统模拟器上进行模拟的前列腺癌患者头四或五天的EPI。我们使用四个参数{M(μi)、Σ(μi)、RMS(μi)、σ(σi)}来表征摆位误差,其中μi和σi是个体患者的均值和标准差,M、Σ和RMS是基础变量(μi和σi)的均值、标准差和均方根。我们通过校正第一天的摆位误差进行了去除系统性成分的模拟。作为比较,我们还对在CT模拟器上模拟的患者以及使用机载千伏级CT成像系统在直线加速器上治疗的患者进行了类似分析,尽管这两个样本中的患者数量有限。我们发现,在左右、前后和头脚方向上,Σ(μi)分别为(2.6, 3.4, 2.4) mm,RMS(σi)分别为(1.5, 1.9, 1.0) mm,这表明系统性误差远大于随机误差。在第一天的测量值与μi之间发现了强相关性,这意味着第一天的测量值是μi的良好预测指标。还计算了第2 - 4天在有和没有第一天校正情况下的相同参数。未校正时,M(μi)2 - 4 =(0.7, 1.6, -1.0) mm,Σ(μi)2 - 4 =(2.6, 3.5, 2.4) mm。校正后,M(μi)2 - 4 =(0.0, 0.4, 0.4) mm,更接近零,Σ(μi)2 - 4 =(1.8, 2.2, 1.2) mm,也更小。虽然使用CT模拟器可以减少系统性误差,但第一天校正的益处仍然可以观察到,尽管幅度较小。因此,如果在第一次分割时在治疗机上对患者进行标记并验证射野,最好使用机载千伏级成像系统,则系统性摆位误差可以显著降低。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验