Loane Edward, Stack Jim, Beatty Stephen, Nolan John M
Macular Pigment Research Group, Department of Chemical & Life Sciences, Waterford Institute of Technology, Waterford, Ireland.
Curr Eye Res. 2007 Jun;32(6):555-64. doi: 10.1080/02713680701418405.
To compare macular pigment optical density using two different heterochromatic flicker photometers.
We measured macular pigment optical density in 121 healthy subjects using heterochromatic flicker photometry.
The mean (+/-SD) macular pigment optical density measured using the Maculometer was 0.394 (+/-0.170), and that using the Densitometer was 0.395 (+/-0.189). The difference in measurements on each instrument was influenced by age and macular pigment levels.
On average, there is no difference in measurements provided by these two instruments. The Maculometer tends to underestimate macular pigment in older subjects and/or those with higher macular pigment compared with the Densitometer.