Suppr超能文献

评估一种新型的黄斑色素光学密度测量仪器:MPS 9000。

An evaluation of a novel instrument for measuring macular pigment optical density: the MPS 9000.

机构信息

Macular Pigment Research Group, Dublin Institute of Technology, Optometry Department, College of Sciences & Health, Dublin 8, Ireland.

出版信息

Acta Ophthalmol. 2012 Mar;90(2):e90-7. doi: 10.1111/j.1755-3768.2011.02294.x. Epub 2011 Nov 8.

Abstract

PURPOSE

Of the antioxidants found in the human retina, only the macular carotenoid quantities can be estimated noninvasively (albeit in a collective fashion), thus facilitating study of their role in that tissue. The aim of this study was to evaluate concordance between macular pigment optical density (MPOD) values recorded on a commercially available instrument, the MPS 9000, with those of an already validated heterochromatic flicker photometry instrument. Also, we assessed and compared test-retest variability for each instrument.

METHODS

Macular pigment optical density at 0.5 retinal eccentricity was measured using two different heterochromatic flicker photometers, the MPS 9000 and the Macular Densitometer(TM), in 39 healthy subjects. Test-retest variability was evaluated separately for each instrument by taking three readings over a 1-week period in 25 subjects.

RESULTS

There was a moderate positive correlation for MPOD at 0.5° of retinal eccentricity between the MPS 9000 and the Macular Densitometer described by the linear equation y = 0.763x + 0.172 (r = 0.68, p < 0.001, r(2) = 0.46); however, a paired-samples t-test showed a significant difference in terms of mean values, with a bias of lower MPOD values being yielded by the MPS 9000 (t = -4.103, p < 0.001). Bland-Altman analysis indicated only moderate agreement between the two instruments, reflected in 95% limits of agreement of 0.1 ± 0.27. Inter-sessional repeatability, expressed as a coefficient of repeatability, ranged from 0.18 to 0.21 [mean (±SD): 0.19 (0.02)] for the MPS 9000 and from 0.11 to 0.12 [mean (±SD): 0.12 (0.01)] for the Macular Densitometer.

CONCLUSION

The results demonstrate that the MPS 9000 consistently yields MPOD readings, which are lower than that found with the Macular Densitometer, and exhibits substantial test-retest variability.

摘要

目的

在人眼视网膜中发现的抗氧化剂中,只有黄斑类胡萝卜素的量可以进行非侵入性(尽管是以集体的方式)估计,从而有利于研究其在该组织中的作用。本研究的目的是评估商业上可用的 MPS 9000 与已经验证的双色闪烁光度计之间黄斑色素光学密度(MPOD)值的一致性。此外,我们评估和比较了每种仪器的测试-重测变异性。

方法

使用两种不同的双色闪烁光度计,即 MPS 9000 和 Macular Densitometer(TM),在 39 名健康受试者中测量 0.5 视网膜偏心处的黄斑色素光学密度。通过在 25 名受试者中在一周内进行三次读数,分别评估每种仪器的测试-重测变异性。

结果

MPS 9000 和 Macular Densitometer 之间在 0.5°视网膜偏心处的 MPOD 具有中等正相关,线性方程为 y = 0.763x + 0.172(r = 0.68,p < 0.001,r² = 0.46);然而,配对样本 t 检验显示平均值存在显著差异,MPS 9000 产生的 MPOD 值较低(t = -4.103,p < 0.001)。Bland-Altman 分析表明两种仪器之间只有中等一致性,95%一致性界限为 0.1 ± 0.27。以可重复性系数表示的间次重复性,MPS 9000 范围为 0.18 至 0.21[平均值(±SD):0.19(0.02)],Macular Densitometer 范围为 0.11 至 0.12[平均值(±SD):0.12(0.01)]。

结论

结果表明,MPS 9000 始终产生的 MPOD 读数低于 Macular Densitometer,并且表现出大量的测试-重测变异性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验