Macular Pigment Research Group, Dublin Institute of Technology, Optometry Department, College of Sciences & Health, Dublin 8, Ireland.
Acta Ophthalmol. 2012 Mar;90(2):e90-7. doi: 10.1111/j.1755-3768.2011.02294.x. Epub 2011 Nov 8.
Of the antioxidants found in the human retina, only the macular carotenoid quantities can be estimated noninvasively (albeit in a collective fashion), thus facilitating study of their role in that tissue. The aim of this study was to evaluate concordance between macular pigment optical density (MPOD) values recorded on a commercially available instrument, the MPS 9000, with those of an already validated heterochromatic flicker photometry instrument. Also, we assessed and compared test-retest variability for each instrument.
Macular pigment optical density at 0.5 retinal eccentricity was measured using two different heterochromatic flicker photometers, the MPS 9000 and the Macular Densitometer(TM), in 39 healthy subjects. Test-retest variability was evaluated separately for each instrument by taking three readings over a 1-week period in 25 subjects.
There was a moderate positive correlation for MPOD at 0.5° of retinal eccentricity between the MPS 9000 and the Macular Densitometer described by the linear equation y = 0.763x + 0.172 (r = 0.68, p < 0.001, r(2) = 0.46); however, a paired-samples t-test showed a significant difference in terms of mean values, with a bias of lower MPOD values being yielded by the MPS 9000 (t = -4.103, p < 0.001). Bland-Altman analysis indicated only moderate agreement between the two instruments, reflected in 95% limits of agreement of 0.1 ± 0.27. Inter-sessional repeatability, expressed as a coefficient of repeatability, ranged from 0.18 to 0.21 [mean (±SD): 0.19 (0.02)] for the MPS 9000 and from 0.11 to 0.12 [mean (±SD): 0.12 (0.01)] for the Macular Densitometer.
The results demonstrate that the MPS 9000 consistently yields MPOD readings, which are lower than that found with the Macular Densitometer, and exhibits substantial test-retest variability.
在人眼视网膜中发现的抗氧化剂中,只有黄斑类胡萝卜素的量可以进行非侵入性(尽管是以集体的方式)估计,从而有利于研究其在该组织中的作用。本研究的目的是评估商业上可用的 MPS 9000 与已经验证的双色闪烁光度计之间黄斑色素光学密度(MPOD)值的一致性。此外,我们评估和比较了每种仪器的测试-重测变异性。
使用两种不同的双色闪烁光度计,即 MPS 9000 和 Macular Densitometer(TM),在 39 名健康受试者中测量 0.5 视网膜偏心处的黄斑色素光学密度。通过在 25 名受试者中在一周内进行三次读数,分别评估每种仪器的测试-重测变异性。
MPS 9000 和 Macular Densitometer 之间在 0.5°视网膜偏心处的 MPOD 具有中等正相关,线性方程为 y = 0.763x + 0.172(r = 0.68,p < 0.001,r² = 0.46);然而,配对样本 t 检验显示平均值存在显著差异,MPS 9000 产生的 MPOD 值较低(t = -4.103,p < 0.001)。Bland-Altman 分析表明两种仪器之间只有中等一致性,95%一致性界限为 0.1 ± 0.27。以可重复性系数表示的间次重复性,MPS 9000 范围为 0.18 至 0.21[平均值(±SD):0.19(0.02)],Macular Densitometer 范围为 0.11 至 0.12[平均值(±SD):0.12(0.01)]。