Color Doppler artifact from metallic carotid clamp.

作者信息

Gooding G A, Saloner D, Eisert W, Nagarkar S

机构信息

Department of Radiology, San Francisco Veterans Administration Medical Center, CA 94121.

出版信息

J Ultrasound Med. 1991 Dec;10(12):691-4. doi: 10.7863/jum.1991.10.12.691.

Abstract

The presence of mirror artifacts in color Doppler has been noted by others. In that report, the artifact arose from scattering at the smooth vessel wall and appeared as signal outside the lumen of the vessel, but with no change in flow direction. As experience increases, recognition of the artifacts of color Doppler will lead to a better understanding and more precise evaluation. This case shows that a band of metal around the carotid artery causes registration errors in color-coded Doppler, and perhaps other metal foreign bodies in the soft tissues have similar potential. The specific appearance of the artifact will depend sensitively on the geometrical configuration of the metal body itself and on its orientation relative to the surrounding anatomy and the ultrasound probe. Appropriate placement of the transducer will reduce such artifact.

摘要

其他人已经注意到彩色多普勒中存在镜面伪像。在那份报告中,伪像是由光滑血管壁的散射引起的,表现为血管腔外的信号,但血流方向没有改变。随着经验的增加,对彩色多普勒伪像的识别将有助于更好地理解和更精确的评估。这个病例表明,颈动脉周围的金属带会导致彩色编码多普勒出现配准误差,软组织中的其他金属异物可能也有类似情况。伪像的具体表现将敏感地取决于金属体本身的几何构型及其相对于周围解剖结构和超声探头的方向。适当放置换能器将减少此类伪像。