Kao Tzu-Jen, Saulnier G J, Xia Hongjun, Tamma Chandana, Newell J C, Isaacson D
Department of Biomedical Engineering, Rensselaer Polytechnic Institute, Troy, NY 12180-3590, USA.
Physiol Meas. 2007 Jul;28(7):S291-9. doi: 10.1088/0967-3334/28/7/S22. Epub 2007 Jun 26.
Electrical impedance tomography (EIT), a non-invasive technique used to image the electrical conductivity and permittivity within a body from measurements taken on the body's surface, could be used as an indicator for breast cancer. Because of the low spatial resolution of EIT, combining it with other modalities may enhance its utility. X-ray mammography, the standard screening technique for breast cancer, is the first choice for that other modality. Here, we describe a radiolucent electrode array that can be attached to the compression plates of a mammography unit enabling EIT and mammography data to be taken simultaneously and in register. The radiolucent electrode array is made by depositing thin layers of metal on a plastic substrate. The structure of the array is presented along with data showing its x-ray absorbance and electrical properties. The data show that the electrode array has satisfactory radiolucency and sufficiently low resistance.