Tang Fuk-hay, Chan Lawrence W C, Lau Hin-pong, Tsui Po-yan, Cheung Chi-wa
Department of Health Technology and Informatics, The Hong Kong Polytechnic University, Hung Hum, Kowloon, Hong Kong.
J Digit Imaging. 2008 Oct;21 Suppl 1(Suppl 1):S113-20. doi: 10.1007/s10278-007-9050-7. Epub 2007 Aug 7.
An intelligence system was used to generate index for scoliosis. Tests were designed to evaluate the consistency of the automatic computer-generated index and to quantify the correlation between Cobb angle and computer generated scoliosis classification index (SCI). A fully automatic computer-generated index can be used to assess the extent of spinal curvature rather than manual measurement on radiographs. This study aims to evaluate the relation of an automatic computer-generated index in assessing the spinal curvature of scoliosis quantitatively on digital chest images. Sixty chest radiographs were obtained in this study. Cobb angle measurement and the index generated were compared by parametric statistical tests. The SCI method was demonstrated to be reproducible. There was also statically significant positive correlation between Cobb angle and SCI (Pearson's correlation: r = 0.9229). The Computer-generated index method is valid and reliable in quantifying measurement of spinal curvature of scoliosis as the correlation between Cobb's angle and SCI in nearly perfect positive for Cobb angle more than 10 degree. It is noted that with widely use of this computer method, this quantitative method proposed is a promising method in improving the reliability of scoliosis assessment and reducing the workload of clinical staff.
使用一种智能系统生成脊柱侧弯指数。设计了测试来评估自动计算机生成指数的一致性,并量化Cobb角与计算机生成的脊柱侧弯分类指数(SCI)之间的相关性。一种全自动计算机生成的指数可用于评估脊柱弯曲程度,而非通过对X光片进行手动测量。本研究旨在评估一种自动计算机生成指数在定量评估数字化胸部图像上脊柱侧弯的脊柱弯曲情况时的关系。本研究获取了60张胸部X光片。通过参数统计检验比较了Cobb角测量值与生成的指数。结果表明SCI方法具有可重复性。Cobb角与SCI之间也存在统计学上显著的正相关(Pearson相关性:r = 0.9229)。对于Cobb角大于10度的情况,由于Cobb角与SCI之间几乎呈完美正相关,计算机生成指数方法在量化脊柱侧弯的脊柱弯曲测量方面是有效且可靠的。值得注意的是,随着这种计算机方法的广泛应用,所提出的这种定量方法在提高脊柱侧弯评估的可靠性和减轻临床工作人员的工作量方面是一种很有前景的方法。