Dennison Christopher R, Wild Peter M, Byrnes Peter W G, Saari Amy, Itshayek Eyal, Wilson Derek C, Zhu Qingan A, Dvorak Marcel F S, Cripton Peter A, Wilson David R
Department of Mechanical Engineering, University of Victoria, P.O. Box 3055, Victoria, B.C., Canada V8W 3P6.
J Biomech. 2008;41(1):221-5. doi: 10.1016/j.jbiomech.2007.07.015. Epub 2007 Aug 29.
Methods were developed to measure intervertebral disc pressure using optical fibre-Bragg gratings (FBGs). The FBG sensor was calibrated for hydrostatic pressure in a purpose-built apparatus and the average sensitivity was determined to be -5.7 +/- 0.085 pm/MPa (mean +/- SD). The average coefficient of determination (r(2)) for the calibration data was 0.99, and the average hysteresis of the sensor was 2.13% of full scale. The FBG was used to measure intradiscal pressure response to compressive load in five lumbar functional spine units. The pressure measured by the FBG sensor varied linearly with applied compressive load with coefficients of determination ranging from 0.84 to 0.97. The FBG sensor's sensitivity to compressive load ranged from 0.702 +/- 0.043 kPa/N (mean +/- SD) in a L1-L2 specimen, to 1.07 +/- 0.069 kPa/N in a L4-L5 specimen. These measurements agree with those of previous studies in lumbar spines. Two strain gauge pressure sensors were also used to measure intradiscal pressure response to compressive load. The measured pressure sensitivity to load ranged from 0.251 kPa/N (L4-L5) to 0.850 kPa/N (L2-L3). The average difference in pressure sensitivity to load between Sensors 1 and 2 was 12.9% of the value for Sensor 1, with a range from 1.1% to 20.4%, which suggests that disc pressure was not purely hydrostatic. This may have contributed to the difference between the responses of the FBG and strain gauge sensors.
已开发出使用光纤布拉格光栅(FBG)测量椎间盘压力的方法。在专门构建的装置中对FBG传感器进行静水压力校准,确定平均灵敏度为-5.7±0.085 pm/MPa(平均值±标准差)。校准数据的平均决定系数(r²)为0.99,传感器的平均滞后为满量程的2.13%。使用FBG测量五个腰椎功能脊柱单元对压缩载荷的椎间盘内压力响应。FBG传感器测量的压力随施加的压缩载荷呈线性变化,决定系数范围为0.84至0.97。FBG传感器对压缩载荷的灵敏度范围从L1-L2标本中的0.702±0.043 kPa/N(平均值±标准差)到L4-L5标本中的1.07±0.069 kPa/N。这些测量结果与先前腰椎研究的结果一致。还使用了两个应变片压力传感器测量对压缩载荷的椎间盘内压力响应。测量的压力对载荷的灵敏度范围从0.251 kPa/N(L4-L5)到0.850 kPa/N(L2-L3)。传感器1和传感器2之间压力对载荷的灵敏度平均差异为传感器1值的12.9%,范围为1.1%至20.4%,这表明椎间盘压力并非纯粹的静水压力。这可能导致了FBG和应变片传感器响应之间的差异。