Hong Samin, Na Kyungdoo, Kim Chan Yun, Seong Gong Je
Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea.
Can J Ophthalmol. 2007 Oct;42(5):707-11. doi: 10.3129/i07-135.
Humphrey Matrix perimetry is a recent development in automated perimetry that uses frequency-doubling technology (FDT). We evaluated the learning effect of Humphrey Matrix perimetry with a full-threshold 30-2 strategy.
Twenty-four patients with primary open-angle glaucoma and early visual field defects as assessed by standard automated perimetry and 24 healthy control subjects participated. All subjects had no prior experience with FDT visual field tests. Humphrey Matrix perimetry using a full-threshold 30-2 strategy was performed 3 times within 1 month. Various visual field indices were compared across multiple tests.
Mean deviation (MD) showed a larger defect at the first test than at the second and third tests (p < 0.001), and the mean total improvement of the MD was 2.80 decibels (dB) in the glaucoma group and 1.71 dB in the control group. Pattern standard deviations (PSDs) were decreased as the tests were repeated, but test duration, fixation losses, and false-positive or false-negative rate were not changed significantly. Additionally, the Glaucoma Hemifield Test (GHT) showed unstable results.
In patients naive to the FDT test, Humphrey Matrix perimetry showed a large learning effect on MD and PSD among the various visual field indices. The GHT results of Humphrey Matrix perimetry showed a large instability.
Humphrey矩阵视野计是自动视野计的一项最新进展,它采用了倍频技术(FDT)。我们用全阈值30-2策略评估了Humphrey矩阵视野计的学习效应。
24例经标准自动视野计评估为原发性开角型青光眼且有早期视野缺损的患者以及24名健康对照者参与研究。所有受试者此前均无FDT视野测试经验。采用全阈值30-2策略的Humphrey矩阵视野计在1个月内进行3次测试。对多次测试的各种视野指标进行比较。
平均偏差(MD)在首次测试时的缺损大于第二次和第三次测试(p<0.001),青光眼组MD的平均总改善为2.80分贝(dB),对照组为1.71dB。模式标准差(PSD)随着测试的重复而降低,但测试时间、固视丢失以及假阳性或假阴性率没有显著变化。此外,青光眼半视野检测(GHT)结果不稳定。
在对FDT测试无经验的患者中,Humphrey矩阵视野计在各种视野指标中对MD和PSD显示出较大的学习效应。Humphrey矩阵视野计的GHT结果显示出较大的不稳定性。