评估髋关节外展和内收力量:更大程度的节段固定能否提高可重复性?

Assessing hip abduction and adduction strength: can greater segmental fixation enhance the reproducibility?

作者信息

Laheru Dhruvkumar, Kerr Jack C, McGregor Alison H

机构信息

Biosurgery and Surgical Technology, Division of Surgery, Oncology and Reproductive Medicine, Faculty of Medicine, Charing Cross Hospital, Imperial College London, UK.

出版信息

Arch Phys Med Rehabil. 2007 Sep;88(9):1147-53. doi: 10.1016/j.apmr.2007.05.017.

Abstract

OBJECTIVE

To improve the reproducibility of testing hip abduction and adduction using an isokinetic dynamometer by a novel testing protocol.

DESIGN

Test-retest design.

SETTING

Biodynamics laboratory.

PARTICIPANTS

Fifteen healthy subjects (9 men, 6 women; age, 22.4+/-0.5 y) were recruited.

INTERVENTIONS

Two setups were compared: the first according to manufacturer's guidelines (setup A) and the second a novel setup incorporating pelvic fixation (setup B). Setups A and B were performed in a random order. Both setups included the same battery of isokinetic (30 degrees/s, 60 degrees/s) and isometric tests, and were repeated 1 week later.

MAIN OUTCOME MEASURES

The peak torque for each abduction and adduction exercise was noted and pelvic motion during testing was recorded.

RESULTS

Setup B significantly (P<.05) reduced transverse pelvic rotation by between 7.5 degrees and 8.0 degrees dependent on test speed. Mean differences for reproducibility of peak torque, ranged from 0.8 to 11.7 Nm. The coefficients of repeatability of both setups were similar, ranging from 21.4 to 56.3 Nm across isokinetic exercises. A similar observation was noted for isometric exercises, with the differences between the coefficients of repeatability ranging from 18.6 to 40.0 Nm.

CONCLUSIONS

Reducing pelvic rotation does not enhance reproducibility of the system and is not related to torque production. Further research is required to determine the optimal test setup.

摘要

目的

通过一种新的测试方案提高使用等速测力计测试髋关节外展和内收的可重复性。

设计

重测设计。

地点

生物动力学实验室。

参与者

招募了15名健康受试者(9名男性,6名女性;年龄,22.4±0.5岁)。

干预措施

比较了两种设置:第一种按照制造商指南(设置A),第二种是采用骨盆固定的新设置(设置B)。设置A和B以随机顺序进行。两种设置都包括相同的等速(30度/秒、60度/秒)和等长测试组,并在1周后重复进行。

主要观察指标

记录每次外展和内收运动的峰值扭矩,并记录测试期间的骨盆运动。

结果

设置B显著(P<0.05)减少了横向骨盆旋转,减少幅度在7.5度至8.0度之间,具体取决于测试速度。峰值扭矩可重复性的平均差异为0.8至11.7牛米。两种设置的重复性系数相似,在等速运动中范围为21.4至56.3牛米。等长运动也有类似观察结果,重复性系数之间的差异范围为18.6至40.0牛米。

结论

减少骨盆旋转并不能提高系统的可重复性,且与扭矩产生无关。需要进一步研究以确定最佳测试设置。