Lu Jin-chun, Xu Hui-ru, Huang Yu-feng
Department of Reproduction & Genetics, PLA Institute of Clinical Laboratory Medicine, Nanjing General Hospital of Nanjing Command, PLA, Nanjing, Jiangsu 210002, China.
Zhonghua Nan Ke Xue. 2007 Sep;13(9):791-4.
To establish a method of determining alpha-glucosidase activity in seminal plasma by semi-automatic biochemistry analyzer.
Alpha-glucosidase activity in seminal plasma from 51 men with normal semen parameters in routine semen analysis were detected by semi-automatic biochemistry analyzer and manual glucose oxidase method, respectively. Then, the intra-assay and inter-assay coefficient variation (CV) and normal reference value were calculated. In the meanwhile, the correlation between the two methods was analyzed.
The intra-assay CVs of 2 seminal plasma samples with different alpha-glucosidase activity detected by semi-automatic biochemistry analyzer were 12.63% and 9.13%, and the inter-assay CVs were 10.67% and 13.49%, respectively. The normal reference value for seminal alpha-glucosidase activity detected with semi-automatic biochemistry analyzer ranged from 102.28 to 555.08 U/L. There was a significantly positive correlation between the semi-automatic biochemistry analyzer and the manual glucose oxidase method (r = 0.792, P < 0.01).
The method of determining alpha-glucosidase activity in seminal plasma by semi-automatic biochemistry analyzer, with its simplicity, less cost of time and reagents, and more reliable result, could be applied to clinical laboratory medicine.
建立一种用半自动生化分析仪测定精浆中α-葡萄糖苷酶活性的方法。
分别采用半自动生化分析仪和手工葡萄糖氧化酶法检测51例常规精液分析中精液参数正常男性的精浆α-葡萄糖苷酶活性。然后,计算批内和批间变异系数(CV)以及正常参考值。同时,分析两种方法之间的相关性。
半自动生化分析仪检测的2份α-葡萄糖苷酶活性不同的精浆样本批内CV分别为12.63%和9.13%,批间CV分别为10.67%和13.49%。半自动生化分析仪检测精浆α-葡萄糖苷酶活性的正常参考值范围为102.28至555.08 U/L。半自动生化分析仪与手工葡萄糖氧化酶法之间存在显著正相关(r = 0.792,P < 0.01)。
用半自动生化分析仪测定精浆中α-葡萄糖苷酶活性的方法,操作简便,省时、省试剂,结果更可靠,可应用于临床检验医学。