使用内镜放置电极进行胃电刺激的可行性。

Feasibility of gastric electrical stimulation by use of endoscopically placed electrodes.

作者信息

Xu Xiaohong, Pasricha Pankaj J, Chen Jiande D Z

机构信息

Division of Gastroenterology, University of Texas Medical Branch, Galveston, Texas, USA.

出版信息

Gastrointest Endosc. 2007 Nov;66(5):981-6. doi: 10.1016/j.gie.2007.05.020.

Abstract

BACKGROUND

Gastric electrical stimulation (GES), which has been reported to have therapeutic potentials for gastroparesis and obesity, involves the surgical placement of electrodes with the patient under general anesthesia. New methods are needed for implanting GES electrodes in a safer and more feasible way.

OBJECTIVE

Our purpose was to investigate the safety and feasibility of placing electrodes endoscopically for GES.

DESIGN AND SETTING

A pilot study.

SUBJECTS

Six female hound dogs that weighed 13 to 22 kg.

INTERVENTIONS

Endoscopically placed electrodes passed through the abdomen and the stomach wall.

MAIN OUTCOME MEASUREMENTS

The study was performed in dogs surgically implanted with gastric serosal electrodes and endoscopically implanted electrodes. The experiment consisted of a 30-minute baseline, a 30-minute GES, and a 30-minute recovery. GES was performed through endoscopically placed electrodes. Gastric slow waves were simultaneously recorded with the serosal electrodes and the endoscopically placed electrodes.

RESULTS

(1) The slow wave frequency recorded from the endoscopically placed electrodes was significantly correlated with that from the serosal electrodes (r = 0.97, P < .002). (2) GES through the endoscopically placed electrodes was able to entrain gastric slow waves. (3) No gastric leakage into the abdominal cavity was noted and the dogs were healthy and comfortable. (4) The endoscopically placed electrodes remained for 2 to 3 weeks.

LIMITATIONS

The fixation of the electrodes needs to be improved for longer-term uses.

CONCLUSIONS

GES may be accomplished without surgery by inserting the electrode wire through the abdomen under endoscopy. The study results indicate that the endoscopically placed electrodes are effective for GES and do not result in any adverse events.

摘要

背景

胃电刺激(GES)已被报道对胃轻瘫和肥胖具有治疗潜力,这需要在全身麻醉下对患者进行电极的手术植入。需要新的方法以更安全、更可行的方式植入GES电极。

目的

我们的目的是研究内镜下放置电极进行胃电刺激的安全性和可行性。

设计与地点

一项初步研究。

对象

6只体重13至22千克的雌性猎犬。

干预措施

通过内镜将电极穿过腹部和胃壁。

主要观察指标

该研究在手术植入胃浆膜电极和内镜植入电极的犬只中进行。实验包括30分钟的基线期、30分钟的胃电刺激期和30分钟的恢复期。通过内镜放置的电极进行胃电刺激。同时用浆膜电极和内镜放置的电极记录胃慢波。

结果

(1)内镜放置电极记录的慢波频率与浆膜电极记录的慢波频率显著相关(r = 0.97,P <.002)。(2)通过内镜放置电极进行胃电刺激能够带动胃慢波。(3)未发现胃内容物漏入腹腔,犬只健康且状态良好。(4)内镜放置的电极保留了2至3周。

局限性

电极的固定需要改进以用于长期使用。

结论

通过在内镜下将电极丝穿过腹部,可不通过手术完成胃电刺激。研究结果表明,内镜放置的电极对胃电刺激有效且不会导致任何不良事件。