Wang Jinnan, Yarnykh Vasily L, Hatsukami Thomas, Chu Baocheng, Balu Niranjan, Yuan Chun
Department of Bioengineering, University of Washington, Seattle, Washington, USA.
Magn Reson Med. 2007 Nov;58(5):973-81. doi: 10.1002/mrm.21385.
In this study, a turbo spin-echo (TSE) based motion-sensitized driven-equilibrium (MSDE) sequence was used as an alternative black-blood (BB) carotid MRI imaging scheme. The MSDE sequence was first optimized for more efficient residual blood signal suppression in the carotid bulb of healthy volunteers. Effective contrast-to-noise ratio (CNR(eff)) and residual signal-to-noise ratio (SNR) in the lumen measured from MSDE images were then compared to those measured from inflow saturation (IS) and double inversion-recovery (DIR) images. Statistically significant higher CNR(eff) and lower lumen SNR were obtained from MSDE images. To assess MSDE sequence in a clinical carotid protocol, 42 locations from six subjects with 50% to 79% carotid stenosis by duplex ultrasound were scanned with both MSDE and multislice DIR. The comparison showed that MSDE images present significantly higher CNR and lower lumen SNR compared to corresponding multislice DIR images. The vessel wall area and mean wall thickness measurements in MSDE images were slightly but significantly lower than those obtained with other blood suppression techniques. In conclusion, in vivo comparisons demonstrated that MSDE sequence can achieve better blood suppression and provide a more accurate depiction of the lumen boundaries by eliminating plaque mimicking artifacts in carotid artery (CA) imaging.
在本研究中,基于涡轮自旋回波(TSE)的运动敏感驱动平衡(MSDE)序列被用作一种替代的黑血(BB)颈动脉MRI成像方案。首先对MSDE序列进行优化,以更有效地抑制健康志愿者颈动脉窦中的残留血液信号。然后将从MSDE图像测量的管腔内有效对比噪声比(CNR(eff))和残留信噪比(SNR)与从流入饱和(IS)和双反转恢复(DIR)图像测量的结果进行比较。从MSDE图像获得的CNR(eff)在统计学上显著更高,管腔SNR更低。为了在临床颈动脉检查方案中评估MSDE序列,对6名经双功超声检查显示颈动脉狭窄50%至79%的受试者的42个部位进行了MSDE和多层DIR扫描。比较结果显示,与相应的多层DIR图像相比,MSDE图像的CNR显著更高,管腔SNR更低。MSDE图像中的血管壁面积和平均壁厚度测量值略低于但显著低于其他血液抑制技术获得的值。总之,体内比较表明,MSDE序列可以实现更好的血液抑制,并通过消除颈动脉(CA)成像中类似斑块的伪影,更准确地描绘管腔边界。