Suppr超能文献

高于美国食品药品监督管理局8T指南的静磁场下人体磁共振成像的安全性:9.4T下的钠成像不影响生命体征或认知能力。

Safety of human MRI at static fields above the FDA 8 T guideline: sodium imaging at 9.4 T does not affect vital signs or cognitive ability.

作者信息

Atkinson Ian C, Renteria Laura, Burd Holly, Pliskin Neil H, Thulborn Keith R

机构信息

Center for Magnetic Resonance Research, University of Illinois at Chicago, Chicago, Illinois 60612, USA.

出版信息

J Magn Reson Imaging. 2007 Nov;26(5):1222-7. doi: 10.1002/jmri.21150.

Abstract

PURPOSE

To assess whether exposure to a 9.4 T static magnetic field during sodium imaging at 105.92 MHz affects human vital signs and cognitive function.

MATERIALS AND METHODS

Measurements of human vital signs and cognitive ability made before and after exposure to a 9.4 T MR scanner and a mock scanner with no magnetic field are compared using a protocol approved by the U.S. Food and Drug Administration (FDA).

RESULTS

Exposure to a 9.4 T static magnetic field during sodium imaging did not result in a statistically significant change in the vital signs or cognitive ability of healthy normal volunteers.

CONCLUSION

Vital sign and cognitive ability measurements made before and after sodium imaging at 9.4 T suggest that performing human MRI at 105.92 MHz in a 9.4 T static magnetic field does not pose a health risk.

摘要

目的

评估在105.92MHz进行钠成像期间暴露于9.4T静磁场是否会影响人体生命体征和认知功能。

材料与方法

使用美国食品药品监督管理局(FDA)批准的方案,比较在暴露于9.4T磁共振成像(MR)扫描仪和无磁场的模拟扫描仪前后对人体生命体征和认知能力的测量结果。

结果

在钠成像期间暴露于9.4T静磁场并未导致健康正常志愿者的生命体征或认知能力发生统计学上的显著变化。

结论

在9.4T进行钠成像前后进行的生命体征和认知能力测量表明,在9.4T静磁场中以105.92MHz进行人体磁共振成像不会带来健康风险。