Suppr超能文献

传统方法与电子根尖定位仪在确定不同乳牙工作长度中的应用:临床研究的系统评价与Meta分析

Conventional methods and electronic apical locator in determining working length in different primary teeth: systematic review and meta-analysis of clinical studies.

作者信息

Mourão Priscila Seixas, Fernandes Izabella Barbosa, Moreira Luana Viviam, Machado Gabrielly Fernandes, Falci Saulo Gabriel Moreira, de Souza Glaciele Maria, Ramos-Jorge Maria Letícia

机构信息

Department of Dentistry, Faculty of Biological and Health Sciences, Federal University of Vales do Jequitinhonha and Mucuri, Diamantina, Brazil.

Department of Oral Health for Children and Adolescents, Faculty of Dentistry, Federal University of Minas Gerais, Belo Horizonte, Brazil.

出版信息

Evid Based Dent. 2025 Jan 8. doi: 10.1038/s41432-024-01105-4.

Abstract

AIM

The aim of the current systematic review was to investigate differences in the measurement of working length in primary teeth between electronic apex locators (EAL) and radiographic methods (conventional radiography [CR] and digital radiography [DR]).

METHODS

A systematic search was conducted in six electronic databases (Medline via PubMed, Web of Science, Science Direct, Virtual Health Library, Scopus and Cochrane Database of Systematic Reviews). Gray literature and reference lists of included studies were also examined. The guiding question of the study was formulated using the PECO strategy (Population, Exposure, Comparator, Outcome): P: primary teeth submitted to pulpectomy; E: measurement of working length using the radiographic methods; C: measurement of working length using electronic apex locator; O: difference in working length. The methodological quality (QUADAS-2) and certainty of the evidence (GRADE) were assessed. Meta-analyses were performed, e foram considerados os tipos de radiografia, grupo de dentes, geração de localizadores apicais. I² statistics were calculated.

RESULTS

Twenty-five studies were included in the review. Of these, eighteen studies were eligible for meta-analysis. Most studies demonstrated fair methodological quality. The working length measured using CR was, on average, 0.52 mm greater when compared to EAL (95% CI: 0.15-0.89, I = 75%). When comparing EAL and DR, no statistically significant difference was found (MD = 0.06 mm, 95% CI: -0.12 to 0.24; I2 = 0%). The majority of articles (n = 20) were considered to have fair quality and five had high quality.

CONCLUSIONS

The findings of the present review demonstrate a difference in the measurement of working length using CR in comparison with EAL. No difference was found between the EAL and DR methods. The certainty of the evidence was considered low for all outcomes.

摘要

目的

本系统评价的目的是研究电子根尖定位仪(EAL)与放射照相方法(传统放射照相[CR]和数字放射照相[DR])在乳牙工作长度测量上的差异。

方法

在六个电子数据库(通过PubMed的Medline、Web of Science、ScienceDirect、虚拟健康图书馆、Scopus和Cochrane系统评价数据库)中进行系统检索。还查阅了灰色文献和纳入研究的参考文献列表。使用PECO策略(人群、暴露、对照、结局)制定研究的指导性问题:P:接受牙髓摘除术的乳牙;E:使用放射照相方法测量工作长度;C:使用电子根尖定位仪测量工作长度;O:工作长度的差异。评估了方法学质量(QUADAS - 2)和证据的确定性(GRADE)。进行了荟萃分析,并考虑了放射照相类型、牙齿组、根尖定位仪一代。计算了I²统计量。

结果

本评价纳入了25项研究。其中,18项研究符合荟萃分析的条件。大多数研究显示出中等的方法学质量。与EAL相比,使用CR测量的工作长度平均长0.52毫米(95%置信区间:0.15 - 0.89,I = 75%)。比较EAL和DR时,未发现统计学上的显著差异(平均差 = 0.06毫米,95%置信区间: - 0.12至0.24;I² = 0%)。大多数文章(n = 20)被认为质量中等,五篇质量高。

结论

本评价的结果表明,与EAL相比,使用CR测量工作长度存在差异。EAL和DR方法之间未发现差异。所有结局的证据确定性均被认为较低。