Shimada Eiji, Pilliar Robert M, Deporter Douglas A, Schroering Robert, Atenafu Eshetu
Department of Dentomaxillofacial Radiology, School of Dentistry, Nihon University, Tokyo, Japan.
Int J Oral Maxillofac Implants. 2007 Nov-Dec;22(6):948-54.
The purpose of this study was to compare patterns of crestal bone remodeling with 2 sintered porous-surfaced dental implant designs during a 14-month functional period.
Two root-form press-fit dental implants were evaluated in healed extraction sites in dog mandibles. The standard (control) design was a press-fit implant with a 2-mm machined collar; the remainder of the implant had a sintered porous surface. The test or "hybrid" design had 3 coronal machined threads instead of a machined collar; the remainder of the implant had a sintered porous surface.
Standardized radiographs indicated significantly less crestal bone loss (0.82 to 0.93 mm versus 1.45 to 1.5 mm) with the hybrid design and a slower approach toward an apparent steady state (12 to 14 months for the hybrid versus 7 months for the standard design). Morphometric assessment of back-scattered scanning electron micrographs confirmed that crestal bone loss was significantly less for the hybrid design on all but the lingual implant aspect.
The addition of coronal threads to an implant relying on a sintered porous surface geometry for its long-term osseointegration reduced the extent of crestal bone loss compared to a machined collar region.