Suppr超能文献

关于浅层千伏放疗中质量保证的必要性

On the need for quality assurance in superficial kilovoltage radiotherapy.

作者信息

Austerlitz C, Mota H, Gay H, Campos D, Allison R, Sibata C

机构信息

The Brody School of Medicine at ECU, Department of Radiation Oncology, Greenville, NC, USA.

出版信息

Radiat Prot Dosimetry. 2008;130(4):476-81. doi: 10.1093/rpd/ncn067. Epub 2008 Mar 6.

Abstract

External auditing of beam output and energy qualities of four therapeutic X-ray machines were performed in three radiation oncology centres in northeastern Brazil. The output and half-value layers (HVLs) were determined using a parallel-plate ionisation chamber and high-purity aluminium foils, respectively. The obtained values of absorbed dose to water and energy qualities were compared with those obtained by the respective institutions. The impact on the prescribed dose was analysed by determining the half-value depth (D(1/2)). The beam outputs presented percent differences ranging from -13 to +25%. The ratio between the HVL in use by the institution and the measurements obtained in this study ranged from 0.75 to 2.33. Such deviations in HVL result in percent differences in dose at D(1/2) ranging from -52 to +8%. It was concluded that dosimetric quality audit programmes in radiation therapy should be expanded to include dermatological radiation therapy and such audits should include HVL verification.

摘要

在巴西东北部的三个放射肿瘤中心,对四台治疗用X射线机的射束输出和能量质量进行了外部审计。分别使用平行板电离室和高纯度铝箔来确定输出和半值层(HVL)。将获得的水吸收剂量值和能量质量与各机构获得的值进行比较。通过确定半值深度(D(1/2))来分析对规定剂量的影响。射束输出的百分比差异范围为-13%至+25%。机构使用的HVL与本研究中获得的测量值之间的比率范围为0.75至2.33。HVL的这种偏差导致D(1/2)处剂量的百分比差异范围为-52%至+8%。得出的结论是,放射治疗中的剂量学质量审计计划应扩大到包括皮肤放射治疗,并且此类审计应包括HVL验证。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验