Tang Sai Chun, Clement Gregory T
Brigham and Women's Hospital, Harvard Medical SChool, Department of Radiology, Boston, MA, USA.
IEEE Trans Ultrason Ferroelectr Freq Control. 2008 Feb;55(2):359-67. doi: 10.1109/TUFFC.2008.654.
The aim of this study is to evaluate the feasibility of using harmonic cancellation for a therapeutic ultrasound transducer excited by a switched-mode power converter without an additional output filter. A switching waveform without the third harmonic was created by cascading two switched-mode power inverter modules at which their output waveforms were pi/3 phase shifted from each other. A PSPICE simulation model for the power converter output stage was developed. The simulated results were in good agreement with the measurement. The waveform and harmonic contents of the acoustic pressure generated by a 1-MHz, self-focused piezoelectric transducer with and without harmonic cancellation have been evaluated. Measured results indicated that the acoustic third harmonicto- fundamental ratio at the focus was small (-48 dB) with harmonic cancellation, compared to that without harmonic cancellation (-20 dB). The measured acoustic levels of the fifth harmonic for both cases with and without harmonic cancellation also were small (-46 dB) compared to the fundamental. This study shows that it is viable to drive a piezoelectric ultrasound transducer using a switched-mode power converter without the requirement of an additional output filter in many high-intensity focused ultrasound (HIFU) applications.
本研究的目的是评估在不使用额外输出滤波器的情况下,利用谐波消除技术对由开关模式电源转换器驱动的治疗超声换能器的可行性。通过级联两个开关模式功率逆变器模块,创建了一个没有三次谐波的开关波形,在这两个模块中,它们的输出波形彼此相差π/3相位。开发了用于功率转换器输出级的PSPICE仿真模型。仿真结果与测量结果吻合良好。评估了具有和不具有谐波消除功能的1MHz自聚焦压电换能器产生的声压的波形和谐波含量。测量结果表明,与没有谐波消除的情况(-20dB)相比,在有谐波消除的情况下,焦点处的声三次谐波与基波之比很小(-48dB)。与基波相比,有和没有谐波消除这两种情况下测量的五次谐波声级也都很小(-46dB)。这项研究表明,在许多高强度聚焦超声(HIFU)应用中,使用开关模式电源转换器驱动压电超声换能器而无需额外的输出滤波器是可行的。