A fluidics comparison of Alcon Infiniti, Bausch & Lomb Stellaris, and Advanced Medical Optics Signature phacoemulsification machines.

作者信息

Georgescu Dan, Kuo Annie F, Kinard Krista I, Olson Randall J

机构信息

Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, 65 Medical Drive, Salt Lake City, UT 84132, USA.

出版信息

Am J Ophthalmol. 2008 Jun;145(6):1014-1017. doi: 10.1016/j.ajo.2008.01.024. Epub 2008 Mar 17.

Abstract

PURPOSE

To compare three phacoemulsification machines for measurement accuracy and postocclusion surge (POS) in human cadaver eyes.

DESIGN

In vitro comparisons of machine accuracy and POS.

METHODS

Tip vacuum and flow were compared with machine indicated vacuum and flow. All machines were placed in two human cadaver eyes and POS was determined.

RESULTS

Vacuum (% of actual) was 101.9% +/- 1.7% for Infiniti (Alcon, Fort Worth, Texas, USA), 93.2% +/- 3.9% for Stellaris (Bausch & Lomb, Rochester, New York, USA), and 107.8% +/- 4.6% for Signature (Advanced Medical Optics, Santa, Ana, California, USA; P < .0001). At 60 ml/minute flow, actual flow and unoccluded flow vacuum (UFV) was 55.8 +/- 0.4 ml/minute and 197.7 +/- 0.7 mm Hg for Infiniti, 53.5 +/- 0.0 ml/minute and 179.8 +/- 0.9 mm Hg for Stellaris, and 58.5 +/- 0.0 ml/minute and 115.1 +/- 2.3 mm Hg for Signature (P < .0001). POS in an 32-year-old eye was 0.33 +/- 0.05 mm for Infiniti, 0.16 +/- 0.06 mm for Stellaris, and 0.13 +/- 0.04 mm for Signature at 550 mm Hg, 60 cm bottle height, 45 ml/minute flow with 19-gauge tips (P < .0001 for Infiniti vs Stellaris and Signature). POS in an 81-year-old eye was 1.51 +/- 0.22 mm for Infiniti, 0.83 +/- 0.06 mm for Stellaris, 0.67 +/- 0.01 mm for Signature at 400 mm Hg vacuum, 70 cm bottle height, 40 ml/minute flow with 19-gauge tips (P < .0001).

CONCLUSIONS

Machine-indicated accuracy, POS, and UFV were statistically significantly different. Signature had the lowest POS and vacuum to maintain flow. Regarding POS, Stellaris was close to Signature; regarding vacuum to maintain flow, Infiniti and Stellaris were similar. Minimizing POS and vacuum to maintain flow potentially are important in avoiding ocular damage and surgical complications.

摘要

目的

比较三台超声乳化仪在人尸体眼中的测量准确性和阻塞后浪涌(POS)。

设计

对仪器准确性和POS进行体外比较。

方法

将探头真空度和流量与仪器显示的真空度和流量进行比较。所有仪器均置于两只人尸体眼中并测定POS。

结果

美国德克萨斯州沃思堡爱尔康公司的Infiniti超声乳化仪的真空度(实际值的百分比)为101.9%±1.7%,美国纽约罗切斯特博士伦公司的Stellaris为93.2%±3.9%,美国加利福尼亚州圣安娜先进医学光学公司的Signature为107.8%±4.6%(P<.0001)。在流速为60毫升/分钟时,Infiniti的实际流速和非阻塞性流速真空度(UFV)分别为55.8±0.4毫升/分钟和197.7±0.7毫米汞柱,Stellaris为53.5±0.0毫升/分钟和179.8±0.9毫米汞柱,Signature为58.5±0.0毫升/分钟和115.1±2.3毫米汞柱(P<.0001)。在550毫米汞柱、瓶高60厘米、流速45毫升/分钟且使用19号探头的情况下,32岁人眼的Infiniti的POS为0.33±0.05毫米,Stellaris为0.16±0.06毫米,Signature为0.13±0.04毫米(Infiniti与Stellaris和Signature相比,P<.0001)。在400毫米汞柱真空度、瓶高70厘米、流速40毫升/分钟且使用19号探头的情况下,81岁人眼的Infiniti的POS为1.51±0.22毫米,Stellaris为0.83±0.06毫米,Signature为0.67±0.01毫米(P<.0001)。

结论

仪器显示的准确性、POS和UFV在统计学上有显著差异。Signature的POS和维持流速所需的真空度最低。在POS方面,Stellaris与Signature相近;在维持流速所需的真空度方面,Infiniti和Stellaris相似。将POS和维持流速所需的真空度降至最低对于避免眼部损伤和手术并发症可能很重要。