Han Young Keun, Miller Kevin M
Department of Ophthalmology, David Geffen School of Medicine at UCLA and the Jules Stein Eye Institute, Los Angeles, California 90095-7002, USA.
J Cataract Refract Surg. 2009 Aug;35(8):1424-9. doi: 10.1016/j.jcrs.2009.03.041.
To compare vacuum rise time, vacuum limit accuracy, and occlusion break surge of 3 new phacoemulsification machines.
Jules Stein Eye Institute and Department of Ophthalmology, David Geffen School of Medicine at UCLA, Los Angeles, California, USA.
The vacuum rise time under normal and enhanced aspiration modes, vacuum limit accuracy, and occlusion break surge of the Infiniti Vision System, Stellaris Vision Enhancement System, and WhiteStar Signature Phacoemulsification System were tested. Vacuum rise time and limit accuracy were measured at limit settings of 400 mm Hg and 600 mm Hg. Surge area was recorded at vacuum limit settings of 200 mm Hg, 300 mm Hg, 400 mm Hg, and 500 mm Hg.
The Infiniti had the fastest vacuum rise times under normal and enhanced aspiration modes. At 4 seconds, the vacuum limit accuracy was greatest with the Infiniti at the 400 mm Hg limit and the Signature at the 600 mm Hg limit. The Stellaris did not reach either vacuum target. The Infiniti performed better than the other 2 machines during testing of occlusion break surge at all vacuum limit settings above 200 mm Hg.
Under controlled laboratory test conditions, the Infiniti had the fastest vacuum rise time, greatest vacuum limit accuracy at 400 mm Hg, and least occlusion break surge. These results can be explained by the lower compliance of the Infiniti system.
比较3款新型超声乳化仪的真空上升时间、真空极限精度和阻塞解除浪涌。
美国加利福尼亚州洛杉矶市加州大学洛杉矶分校大卫·格芬医学院朱尔斯·斯坦眼科研究所及眼科。
对Infiniti视觉系统、Stellaris视觉增强系统和WhiteStar Signature超声乳化系统在正常和增强抽吸模式下的真空上升时间、真空极限精度和阻塞解除浪涌进行测试。在400 mmHg和600 mmHg的极限设置下测量真空上升时间和极限精度。在200 mmHg、300 mmHg、400 mmHg和500 mmHg的真空极限设置下记录浪涌面积。
Infiniti在正常和增强抽吸模式下具有最快的真空上升时间。在4秒时,Infiniti在400 mmHg极限下的真空极限精度最高,Signature在600 mmHg极限下的真空极限精度最高。Stellaris未达到任何一个真空目标。在高于200 mmHg的所有真空极限设置下进行阻塞解除浪涌测试时,Infiniti的表现优于其他两款仪器。
在受控的实验室测试条件下,Infiniti具有最快的真空上升时间、400 mmHg时最高的真空极限精度以及最小的阻塞解除浪涌。这些结果可以通过Infiniti系统较低的顺应性来解释。