Objective measurement of postocclusion surge during phacoemulsification in human eye-bank eyes.

作者信息

Georgescu Dan, Payne Marielle, Olson Randall J

机构信息

Department of Ophthalmology, John A. Moran Eye Center, Salt Lake City, Utah 84132, USA.

出版信息

Am J Ophthalmol. 2007 Mar;143(3):437-40. doi: 10.1016/j.ajo.2006.11.017. Epub 2006 Dec 18.

Abstract

PURPOSE

To objectively compare the postocclusion vacuum surge among different phacoemulsification machines and devices.

DESIGN

Experimental study.

METHODS

Infiniti, Legacy, Millennium, and Sovereign were tested in an eye-bank eye. All the machines were tested with 20-gauge non-ABS tips, 430 mm Hg vacuum pressure, 24 ml/minute aspiration rate, peristaltic pump, and 75 cm bottle height. In addition, Infiniti and Legacy were also tested with 20-gauge bypass tips (ABS), 125 cm bottle height, and 40 ml/minute flow rate. We also tested 19-gauge tips with Infiniti and Sovereign and the venturi pump for Millennium.

RESULTS

Significant differences were found between all the machines tested with Millennium peristaltic generating the least and Millennium Venturi the most surge. ABS tips significantly decreased the surge for Legacy but not for Infiniti. Cruise Control (CC) had a significant effect on Sovereign but not on Millennium. Increasing the bottle height decreased surge while increasing the flow increased surge for both Infiniti and Legacy. The 19-gauge tips increased surge for both Infiniti and Sovereign.

CONCLUSIONS

Surge varied over a range of 40 microm to more than 2 mm. ABS and CC decrease surge, especially when the machine is not functioning near the limits of surge prevention. Certain parameters, such as a 19-gauge tip and high flow, dramatically increased surge, whereas elevating the bottle ameliorates it. Understanding the impact of all these features will help in minimizing the problem.

摘要

目的

客观比较不同超声乳化仪及设备的阻塞后真空波动情况。

设计

实验研究。

方法

在眼库眼球上对Infiniti、Legacy、Millennium和Sovereign进行测试。所有仪器均使用20号非ABS穿刺头、430毫米汞柱真空压力、24毫升/分钟抽吸速率、蠕动泵以及75厘米瓶高进行测试。此外,Infiniti和Legacy还使用20号旁路穿刺头(ABS)、125厘米瓶高以及40毫升/分钟流速进行测试。我们还使用Infiniti和Sovereign测试了19号穿刺头,并对Millennium使用了文丘里泵进行测试。

结果

在所有测试仪器中发现了显著差异,使用蠕动泵的Millennium产生的波动最小,而使用文丘里泵的Millennium产生的波动最大。ABS穿刺头显著降低了Legacy的波动,但对Infiniti没有效果。巡航控制(CC)对Sovereign有显著影响,但对Millennium没有影响。增加瓶高可降低波动,而增加流速则会增加Infiniti和Legacy的波动。19号穿刺头增加了Infiniti和Sovereign的波动。

结论

波动范围在40微米至超过2毫米之间。ABS和CC可降低波动,尤其是当仪器在接近防波动极限运行时。某些参数,如19号穿刺头和高流速,会显著增加波动,而提高瓶高则可改善波动。了解所有这些特性的影响将有助于将问题最小化。