瓶高和抽吸速率对英菲尼迪和千禧蠕动式超声乳化仪阻塞后压力波动的影响。

Effect of bottle height and aspiration rate on postocclusion surge in Infiniti and Millennium peristaltic phacoemulsification machines.

作者信息

Ward Matthew S, Georgescu Dan, Olson Randall J

机构信息

Department of Ophthalmology and Visual Sciences, Moran Eye Center, University of Utah, Salt Lake City, UT 84132, USA.

出版信息

J Cataract Refract Surg. 2008 Aug;34(8):1400-2. doi: 10.1016/j.jcrs.2008.04.042.

Abstract

PURPOSE

To assess how flow and bottle height affect postocclusion surge in the Infiniti (Alcon, Inc.) and Millennium (Bausch & Lomb) peristaltic machines.

SETTING

John A. Moran Eye Center Clinical Laboratories, University of Utah, Salt Lake City, Utah.

METHODS

Postocclusion anterior chamber depth changes were measured in human eye-bank eyes using A-scan. Surge was simulated by clamping the aspiration tubing and releasing it at maximum vacuum. In both machines, surge was measured (1) with aspiration held constant at 12 mL/min and bottle heights at 60, 120, and 180 cm and (2) with bottle height held constant at 60 cm and aspiration rates at 12, 24, and 36 mL/min.

RESULTS

Surge decreased approximately 40% with each 60 cm increase in bottle height in the Infiniti. It was constant at all bottle heights in the Millennium. At 12 and 24 mL/min aspiration rates, surge in the Millennium was less than half that in the Infiniti (P<.001).

CONCLUSIONS

Postocclusion surge decreased linearly with increasing bottle height in the Infiniti system and was relatively constant with increasing bottle height in the Millennium system. The Millennium may offer a more stable phacoemulsification platform with respect to surge at a higher aspiration rate.

摘要

目的

评估流速和瓶高如何影响Infiniti(爱尔康公司)和Millennium(博士伦公司)蠕动式超声乳化仪的阻塞后浪涌。

设置

犹他大学盐湖城分校约翰·A·莫兰眼科中心临床实验室。

方法

使用A超测量人眼库眼球的阻塞后前房深度变化。通过夹住抽吸管道并在最大真空度下释放来模拟浪涌。在两台仪器中,测量浪涌的情况如下:(1) 抽吸速率保持在12 mL/分钟不变,瓶高分别为60、120和180厘米;(2) 瓶高保持在60厘米不变,抽吸速率分别为12、24和36 mL/分钟。

结果

在Infiniti中瓶高每增加60厘米,浪涌大约降低40%。在Millennium中,浪涌在所有瓶高下保持恒定。在抽吸速率为12和24 mL/分钟时,Millennium中的浪涌不到Infiniti中的一半(P<0.001)。

结论

在Infiniti系统中,阻塞后浪涌随瓶高增加呈线性降低,而在Millennium系统中,浪涌随瓶高增加相对恒定。就较高抽吸速率下的浪涌而言,Millennium可能提供一个更稳定的超声乳化平台。